Specifications (Continued)
Interface
IEEE-488.2 Interface
Features:
Reading Rate:
Software Support:
Serial Interface
Electrical Format:
Max. Baud Rate:
Connector:
Reading Rate:
Special Interface Features:
Alarms
Number:
Data Source:
Settings:
Actuators:
Relays
Number:
Contacts:
Contact Rating:
Operation:
Connector:
Analog Voltage Output (when not used as control loop 2 output)
Scale:
Update Rate:
Data Source:
Settings:
Range:
Resolution:
Accuracy:
Maximum Output Power:
Minimum Load Resistance:
Source Impedance:
General
Ambient Temperature:
Power Requirement:
Size:
Weight:
Ordering Information
Standard Temperature Controllers
Part Number
332S
332S-T1
332S-T2
Power Options
(Select one, the instrument will be configured for selected power and fuses)
VAC-100
VAC-120
VAC-220
VAC-240
VAC-120-All
1-10
Lake Shore Model 332 Temperature Controller User's Manual
SH1,AH1,T5,L4,SR1,RL1,PP0,DC1,DT0,C0,E1
To 10 readings per second on each input
LabView Driver
RS-232C
9600 Baud
DE-9
To 10 readings per second on each input (at 9600 Baud)
Model 330 command emulation mode
4, High and Low for each input
Temperature, Sensor Units, Linear Equation
Source, High & Low Setpoint, Deadband, Latching or Non-Latching, Audible On/Off
Display annunciator, beeper, relays
2
Normally Open (NO), Normally Closed (NC), and Common (COM)
30 VDC at 5 A
Activate relays on high, low, or both alarms for either input or manual
Detachable terminal block
User selected
10 readings per second
Temperature, Sensor Units, Linear Equation
Input, Source, Top of scale, Bottom of scale, or manual
±10 V
0.3 mV
±2.5 mV
1 W (jumper selected)
100 Ω (short-circuit protected)
0.01 Ω
15 – 35 °C at rated accuracy. 10 – 40 °C at reduced accuracy
100, 120, 220, 240 VAC, +6% –10%, 50 or 60 Hz., 150 VA
217 mm wide × 90 mm high × 368 mm deep (8.5 × 3.5 × 14.5 inches), half rack
4.8 kilograms (10.5 pounds)
Description (Input configuration cannot be changed in the field)
Two Diode/Resistor Inputs
One Diode/Resistor, One Thermocouple Input
Two Thermocouple Inputs
Instrument configured for 100 VAC with U.S. power cord
Instrument configured for 120 VAC with U.S. power cord
Instrument configured for 220 VAC with European power cord
instrument configured for 240 VAC with European power cord
Instrument configured for 120 VAC with U.S. power cord and universal European
power cord and fuses for 220/240 setting (extra charge for this option)