BEEP
Alarm Beeper Command
Input:
BEEP <state>[term]
n
Format:
<state>
Remarks: Enables or disables system beeper sound when an alarm condition is met.
BEEP?
Alarm Beeper Query
Input:
BEEP?
Returned: <state> [term]
n
Format:
BRIGT
Display Brightness Command
Input:
BRIGT <bright>[term]
n
Format:
<bright>
BRIGT?
Display Brightness Query
Input:
BRIGT?[term]
Returned: <bright>[term]
n
Format:
CMODE
Control Loop Mode Command
Input:
CMODE <loop>, <mode>[term]
n,n
Format:
<loop>
<mode>
Example: CMODE 1,4[term] – Control Loop 1 uses PID AutoTuning.
CMODE?
Control Loop Mode Query
Input:
CMODE? <loop>[term]
Format: n
<loop>
Returned: <mode>[term]
Format: n
CRDG?
Celsius Reading Query
Input:
CRDG? <input>[term]
Format: a
<input>
Returned: <temp value>[term]
Format: ±nnnnnn
Remarks: Also see the RDGST? command.
6-28
Lake Shore Model 332 Temperature Controller User's Manual
0 = Off, 1 = On.
(Refer to command for description)
0 = 25%, 1 = 50%, 2 = 75%, 3 = 100%. Default = 2.
(Refer to command for description)
Specifies which loop to configure: 1 or 2.
Specifies the control mode. Valid entries:
3 = Open Loop, 4 = AutoTune PID, 5 = AutoTune PI, 6 = AutoTune P.
Specifies which loop to query: 1 or 2.
(Refer to command for description)
A or B
1 = Manual PID, 2 = Zone,
Remote Operation