HTR?
Heater Output Query
Input:
HTR?[term]
Returned: <heater value>[term]
+nnn.n
Format:
<heater value>
HTRST?
Heater Status Query
Input:
HTRST?[term]
Returned: <error code>[term]
n
Format:
<error code>
IEEE
IEEE-488 Interface Parameter Command
Input:
IEEE <terminator>, <EOI enable>, <address>[term]
n,n,nn
Format:
<terminator>
<EOI enable> Sets EOI mode: 0 = enabled, 1 = disabled.
<address>
Example: IEEE 0,0,4[term] – After receipt of the current terminator, the instrument uses EOI mode,
uses <CR><LF> as the new terminator, and responds to address 4.
IEEE?
IEEE-488 Interface Parameter Query
Input:
IEEE?[term]
Returned: <terminator>, <EOI enable>, <address>[term]
n,n,nn
Format:
INCRV
Input Curve Number Command
Input:
INCRV <input>, <curve number>[term]
a,nn
Format:
<input>
<curve number> Specifies which curve the input uses. If specified curve parameters do
Remarks: Specifies the curve an input uses for temperature conversion.
Example: INCRV A,23[term] – Input A uses User Curve 23 for temperature conversion.
INCRV?
Input Curve Number Query
Input:
INCRV? <input>[term]
Format: a
<input>
Returned: <curve number>[term]
Format: nn
6-32
Lake Shore Model 332 Temperature Controller User's Manual
Loop 1 heater output in percent (%). Use AOUT? for Loop 2.
Heater error code: 0 = no error, 1 = heater open load, 2 = heater short.
Specifies the terminator. Valid entries: 0 = <CR><LF>,1 = <LF><CR>,
2 = <LF>, 3 = no terminator (must have EOI enabled).
Specifies the IEEE address: 1 – 30. (Address 0 and 31 are reserved.)
(Refer to command for description)
Specifies which input to configure: A or B.
not match the input, the curve number defaults to 0. Valid entries:
0 = none, 1 – 20 = standard curves, 21 – 41 = user curves.
Specifies which input to query: A or B.
(Refer to command for description)