Lakeshore 332 User Manual page 133

Temperature controller
Table of Contents

Advertisement

LOCK?
Front Panel Keyboard Lock Query
Input:
LOCK?[term]
Returned: <state>, <code>[term]
n,nnn
Format:
MDAT?
Minimum/Maximum Data Query
Input:
MDAT? <input>[term]
Format: a
<input>
Returned: <min value>,<max value>[term]
Format: ±nnnnnn,±nnnnnn
Remarks: Returns the minimum and maximum input data. Also see the RDGST? command.
MNMX
Minimum and Maximum Input Function Parameter Command
Input:
MNMX <input>, <source>[term]
a,n
Format:
<input>
<source>
Example: MNMX B,3[term] – Input B min/max function is on and processes data from the input sensor
units reading.
MNMX?
Minimum and Maximum Input Function Parameter Query
Input:
MNMX? <input>[term]
Format: a
<input>
Returned: <source>[term]
Format: n
MNMXRST
Minimum and Maximum Function Reset Command
Input:
MNMXRST[term]
Remarks: Resets the minimum and maximum data for all inputs.
MODE
Remote Interface Mode Command
Input:
MODE <mode>[term]
n
Format:
<mode>
Example: MODE 2[term] – Places the Model 332 into remote mode with local lockout.
Remote Operation
Lake Shore Model 332 Temperature Controller User's Manual
(Refer to command for description)
Specifies which input to query: A or B.
Specifies input to configure: A or B.
Specifies input data to process through max/min. Valid entries: 1 = kelvin,
2 = Celsius, 3 = sensor units, 4 = linear data.
Specifies which input to query: A or B.
(Refer to command for description)
0 = local, 1 = remote, 2 = remote with local lockout.
6-35