Date; Init - Keithley 2700 Service Manual

Data acquisition system
Hide thumbs Also See for 2700:
Table of Contents

Advertisement

B-18
Calibration Reference

:DATE?

(:CALibration:PROTected:CARD1:DATE?)
(:CALibration:PROTected:CARD2:DATE?)
Purpose
Format
Response
Description
NOTE
Example

:INIT

(:CALibration:PROTected:CARD1:INITiate)
Purpose
Format
Parameter
Description
Example
To request the Model 7700 calibration date.
:cal:prot:card1:date?
:cal:prot:card2:date?
<year>, <month>, <day>
The :CARD1:DATE? and :CARD2:DATE? queries allow you to read
back the calibration date from a Model 7700 in Slot 1 and Slot 2
respectively.
The card calibration date is automatically set to the Model 2700 real time clock date
when the card is calibrated.
:CAL:PROT:CARD1:DATE?
To initiate Model 7700 calibration procedures.
:cal:prot:card1:init
None
The :INIT command enables Model 7700 calibration when performing
these procedures over the bus. This command must be sent to the unit after
sending the :CODE command, but before performing Model 7700
calibration.
:CAL:PROT:CARD1:INIT
Request card 1 cal date.
Initiate 7700 calibration.

Advertisement