2-16
Calibration
DC current calibration
After the 1MΩ resistance point has been calibrated, follow these steps for DC current
calibration:
1.
2.
NOTE
Table 2-10
DC current calibration programming steps
Calibration
step
10mA
100mA
1A
*Change parameter if using different current.
AC voltage calibration
Follow these steps for AC voltage calibration:
1.
2.
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2700 as
shown in
Figure
2-3.
Perform the calibration steps listed in
Set the calibrator to the indicated current, and make sure the unit is in operate.
(Use the recommended current if possible.)
Send the indicated programming command. (Change the current parameter if you
are using a different calibration current.)
Wait until the Model 2700 completes each step before continuing.
If you are performing DC-only calibration, proceed to "Programming calibration
dates."
Calibrator
current
10.00000mA
100.00000mA
1.000000A
Connect the calibrator to the Model 2700 INPUT HI and LO terminals as shown in
Figure
2-4.
Perform the calibration steps summarized in
Set the calibrator to the indicated voltage and frequency, and make sure the unit is
in operate. (You must use the stated voltage and frequency.)
Send the indicated programming command.
Wait until the Model 2700 completes each step before continuing.
Table
2-10. For each step:
Calibration command*
:CAL:PROT:DC:STEP10 10E-3
:CAL:PROT:DC:STEP11 100E-3
:CAL:PROT:DC:STEP12 1
Table
2-11. For each step:
Parameter range
9E-3 to 11E-3
90E-3 to 110E-3
0.9 to 1.1