2-16
Calibration
DC current calibration
After the 1MΩ resistance point has been calibrated, follow these steps for DC current
calibration:
1.
2.
NOTE
Table 2-10
DC current calibration programming steps
Calibration
step
10mA
100mA
1A
*Change parameter if using different current.
AC voltage calibration
Follow these steps for AC voltage calibration:
1.
2.
Connect the calibrator to the AMPS and INPUT LO terminals of the Model 2700 as
shown in
Figure
2-3.
Perform the calibration steps listed in
•
Set the calibrator to the indicated current, and make sure the unit is in operate.
(Use the recommended current if possible.)
•
Send the indicated programming command. (Change the current parameter if you
are using a different calibration current.)
•
Wait until the Model 2700 completes each step before continuing.
If you are performing DC-only calibration, proceed to "Programming calibration
dates."
Calibrator
current
10.00000mA
100.00000mA
1.000000A
Connect the calibrator to the Model 2700 INPUT HI and LO terminals as shown in
Figure
2-4.
Perform the calibration steps summarized in
•
Set the calibrator to the indicated voltage and frequency, and make sure the unit is
in operate. (You must use the stated voltage and frequency.)
•
Send the indicated programming command.
•
Wait until the Model 2700 completes each step before continuing.
Table
2-10. For each step:
Calibration command*
:CAL:PROT:DC:STEP10 10E-3
:CAL:PROT:DC:STEP11 100E-3
:CAL:PROT:DC:STEP12 1
Table
2-11. For each step:
Parameter range
9E-3 to 11E-3
90E-3 to 110E-3
0.9 to 1.1