Table 1-11 Plug-In Module Dci Limits - Keithley 2700 Service Manual

Data acquisition system
Hide thumbs Also See for 2700:
Table of Contents

Advertisement

2.
Install the Model 7700 in Slot 1 of the Model 2700, then turn on the power, and allow
the unit to warm up for two hours before proceeding. Be sure the front panel INPUTS
switch is set to the REAR position.
3.
Select the DC current measurement function by pressing the DCI key.
4.
Set the Model 2700 for the 20mA range. Close Channel 21 by pressing the CLOSE key,
and keying in 121.
5.
Source positive and negative full-scale currents for each of the ranges listed in
Table
1-11, and verify that the readings for each range are within stated limits.
6.
Press the OPEN key to open Channel 21.
Table 1-11
Plug-in module DCI limits
DCI range
Applied DC current*
20mA
20.0000mA
100mA
100.0000mA
1A
1.000000A
3A
3.000000A**
* Source positive and negative currents with values shown.
** If the Fluke 5725 amplifier is not available, apply 2.2A from calibrator. Reading limits for 2.2A
input are: 2.197240 to 2.202760A.
Performance Verification
Reading limits (1 year, 18°C to 28°C)
19.89960 to 20.01040mA
99.9100 to 100.0900mA
0.999160 to 1.000840A
2.99628 to 3.00372A
1-23