Keithley 2700 Service Manual page 128

Data acquisition system
Hide thumbs Also See for 2700:
Table of Contents

Advertisement

A-8
Specifications
7700 20-Channel Differential Multiplexer w/Automatic CJC
GENERAL
20 CHANNELS: 20 channels of 2-pole relay input. All channels config-
urable to 4-pole.
2 CHANNELS: 2 channels of current only input.
RELAY TYPE: Latching electromechanical.
ACTUATION TIME: <3ms.
CAPABILITIES
CHANNELS 1-20: Multiplex one of 20 2-pole or one of 10 4-pole signals
into DMM.
CHANNELS 21-22: Multiplex one of 2 2-pole current signals into
DMM.
INPUTS
MAXIMUM SIGNAL LEVEL:
Channels (1-20): 300V DC or rms, 1A switched, 60W, 125VA maxi-
mum.
Channels (21-22): 60V DC or 30V rms, 3A switched, 60W, 125VA
maximum.
CONTACT LIFE (typ): >10
5
operations at max signal level.
>10
8
operations cold switching.
CONTACT RESISTANCE: <1Ω at end of contact life.
CONTACT POTENTIAL: <±500nV typical per contact, 1µV max.
<±500nV typical per contact pair, 1µV max.
OFFSET CURRENT: <100pA.
CONNECTOR TYPE: Screw terminal, #20 AWG wire size.
ISOLATION BETWEEN ANY TWO TERMINALS: >10
ISOLATION BETWEEN ANY TERMINAL AND EARTH: >10
CROSS TALK (10MHz, 50Ω Load): <–40dB.
INSERTION LOSS (50Ω Source, 50Ω Load): <0.1dB below 1MHz.
COMMON MODE VOLTAGE: 300V between any terminal and chassis.
T/C COLD JUNCTION: 1.0°C (18°–28°C Mainframe Temp)
1.5°C (0°–18°C & 28°–50°C Mainframe Temp).
ENVIRONMENTAL:
OPERATING ENVIRONMENT: Specified for 0°C to 50°C.
Specified to 80% R.H. at 35°C.
STORAGE ENVIRONMENT: –25°C to 65°C.
WEIGHT: 0.45kg (1 lb).
Ω, <100pF.
10
Ω, <200pF.
9
<3dB below 2MHz.
HI
Card Input
LO
HI
Card Sense
LO
Cold junction
Ref x3
HI
Channel 1
LO
(Channels 2–9)
HI
Channel 10
LO
Channel 23
2-Pole (Open)
4-Pole (Closed)
(see Note)
Cold junction
Ref x3
HI
Channel 11
LO
(Channels 12–19)
HI
Channel 20
LO
3A
HI
Channel 21
LO
3A
HI
Channel 22
LO
AMPS
Card
LO
Channel 25
(see Note)
Backplane
isolation
HI
Input
LO
Channel 24
(see Note)
Backplane
isolation
HI
Sense
To Model 2700
LO
Backplane
AMPS
NOTE
Channels 23–25 in this schematic
refer to the designations used for
control and not actual available channels.
For more information, refer to the
ROUTe:MULTiple command section
in the Model 2700 User's Manual.
Channels 24 and 25 can be individually
controlled using ROUTe:MULTiple if the
module is not to be connected to the
internal DMM.

Advertisement

Table of Contents
loading

Table of Contents