Step8; Step9; Step10 - Keithley 2700 Service Manual

Data acquisition system
Hide thumbs Also See for 2700:
Table of Contents

Advertisement

:STEP8

(:CALibration:PROTected:DC:STEP8)
Purpose
Format
Parameter
Description
Example

:STEP9

(CALibration:PROTected:DC:STEP9)
Purpose
Format
Parameter
Description
Example

:STEP10

(CALibration:PROTected:DC:STEP10)
Purpose
Format
Parameter
Description
Example
To program the 100kΩ 4-wire comprehensive calibration step.
:cal:prot:dc:step8 <Cal_resistance>
<Cal_resistance> = 90E3 to 110E3 [Ω]
:STEP8 programs the 100kΩ 4-wire resistance comprehensive calibration
step. The allowable range of the calibration resistance parameter is from
90E3 to 110E3, but 100E3 is recommended for best results.
:CAL:PROT:DC:STEP8 100E3
To program the 1MΩ comprehensive calibration step.
:cal:prot:dc:step9 <Cal_resistance>
<Cal_resistance> = 900E3 to 1.1E6 [Ω]
:STEP9 programs the 1MΩ comprehensive calibration step. The allowable
range of the calibration resistance parameter is from 900E3 to 1.1E6. Use
the 1E6 value whenever possible, or the closest possible value.
:CAL:PROT:DC:STEP9
To program the 10mA comprehensive calibration step.
:cal:prot:dc:step10 <Cal_current>
<Cal_current> = 9E-3 to 11E-3 [A]
:STEP10 programs the 10mA comprehensive calibration step. The allow-
able range of the calibration current parameter is from 9E-3 to 11E-3. Use
the 10E-3 value whenever possible for best results.
:CAL:PROT:DC:STEP10 10E-3
Calibration Reference
Program 100kΩ step.
Program 1MΩ calibration step.
Program 10mA step.