Front Panel Manufacturing Calibration; Figure 2-5 Function Generator Connections For Manufacturing Calibration - Keithley 2700 Service Manual

Data acquisition system
Hide thumbs Also See for 2700:
Table of Contents

Advertisement

2-20
Calibration

Front panel manufacturing calibration

1.
2.
3.
4.
5.
6.
7.
Figure 2-5
Function generator connections for manufacturing calibration
Model 2700
Model 2700 Multimeter / Data Acquisition System
RATIO
CH AVG
CONT
OCOMP
MATH O U T P U T
Ω2
Ω4
DCV
ACV
ACI
SHIFT
DCI
DELAY
HOLD
LIMITS
ON/OFF
TYPE
MONITOR
LOCAL
EX TRIG
TRIG
STORE
FILTER
REL
RECALL
POWER
SAVE
SETUP
CONFIG
HALT
TEST
LSYNC
OPEN
CLOSE
STEP
SCAN
DIGITS RATE
8.
9.
Install the shorted calibration extender board (see "Extender board preparation" earlier
in this section) in scanner card Slot 1, and select the rear inputs with the INPUTS
switch. Allow three minutes for thermal equilibrium.
Press in and hold the OPEN key while turning on the power.
Press SHIFT then TEST. then display CALIB: TEST with the up or down range key.
Press ENTER, select RUN, then enter the appropriate calibration code (default:
002700).
Select ALL at the CAL:RUN prompt.
Press ENTER to perform the first manufacturing calibration step.
Perform the entire front panel comprehensive calibration procedure discussed earlier in
this section. (See "Comprehensive calibration" earlier in this section.)
Connect the function generator to the Model 2700 front panel INPUT jacks as shown in
Figure
2-5. Select the front input jacks with the INPUTS switch.
BNC-to-Dual
Banana Plug
Adapter
Integra Series
SENSE
INPUT
Ω 4 WIRE
HI
350V
1000V
!
PEAK
PEAK
LO
500V
PERIOD SENSOR
PEAK
INPUTS
FREQ
TEMP
RANGE
F
R
CH-OFF
CARD
AUTO
FRONT/REAR
3A 250V
AMPS
RANGE
GPIB
RS-232
EXIT
ENTER
After the last AC current calibration step, the instrument will prompt you to enter 3Hz
at 1V RMS and 1kHz with the following prompts:
Low-frequency cal: Set the function generator to output a 1V RMS, 3Hz sine
wave. Use the left and right arrow keys, and the range keys to adjust the display to
agree with the generator amplitude you measured previously, then press ENTER.
Frequency cal: Set the function generator to output a 1V RMS, 1kHz sine wave.
Enter 1.000000kHz at the prompt, then press ENTER.
Set the calibration dates, then save calibration to complete the process.