Ac Current - Keithley 2700 Service Manual

Data acquisition system
Hide thumbs Also See for 2700:
Table of Contents

Advertisement

AC SPECIFICATIONS
Function
Range
Resolution
Voltage
2
100.0000 mV
1.000000
V
10.00000
V
100.0000
V
750.000
V
Current
2
1.000000
A
3.00000
A
Frequency
4
100 mV
0.333 ppm
and Period
to
3.33 ppm
750
V
33.3 ppm
Additional Uncertainty ±(% of reading)
Low Frequency Uncertainty
20 Hz - 30 Hz
30 Hz - 50 Hz
50 Hz - 100 Hz
100 Hz - 200 Hz
200 Hz - 300 Hz
>300 Hz
CREST FACTOR:
5
1 -2
Additional Uncertainty:
0.05
AC MEASUREMENT CHARACTERISTICS
AC Volts
MEASUREMENT METHOD: AC-coupled, True RMS.
INPUT IMPEDANCE: 1MΩ ±2% // by <100pF.
INPUT PROTECTION: 1000Vp or 400VDC. 300Vrms with plug in modules.

AC Current

MEASUREMENT METHOD: AC-coupled, True RMS.
SHUNT RESISTANCE: 0.1Ω.
BURDEN VOLTAGE: 1A <0.3Vrms, 3A <1Vrms. Add 1Vrms when used with plug in modules.
INPUT PROTECTION: 3A, 250V fuse.
Frequency and Period
MEASUREMENT METHOD: Reciprocal Counting technique.
GATE TIME: SLOW 1s, MED 100ms, and FAST 10ms.
AC General
AC CMRR
6
: 70dB.
MAXIMUM CREST FACTOR: 5 at full-scale.
VOLT HERTZ PRODUCT: <= 8 × 10
1
Calibration
3 Hz–
Cycle
10 Hz
0.1 µV
90 Days
0.35 + 0.03
1.0 µV
10 µV
1 Year
0.35 + 0.03
100 µV
1.0 µV
(Temp. Coeff.)/°C
3
0.035 + .003
3 Hz–10 Hz
1.0 µA
90 Day/1 Yr.
0.30 + 0.04
10 µA
0.35 + 0.06
(Temp. Coeff.)/°C
3
0.035 + 0.006
(3 Hz–500 kHz) (333 ms–2 µs)
90 Day/ 1 Yr.
100 ppm + 0.333 ppm (SLOW, 1s gate)
100 ppm + 3.33 ppm (MED, 100ms gate)
100 ppm + 33.3 ppm (FAST, 10ms gate)
MED
FAST
0.3
0
0
1.0
0
0.18
0
0.10
0
0
2 - 3
3 - 4
4 - 5
0.15
0.30
0.40
7
.
Accuracy: ±(% of reading + % of range), 23°C ± 5°C
10 Hz–
20 kHz–
20 kHz
50 kHz
0.05 + 0.03
0.11 + 0.05
0.06 + 0.03
0.12 + 0.05
0.005 + .003
0.006 + .005
10 Hz–5 kHz
0.10 + 0.04
0.15 + 0.06
0.015 + 0.006
Specifications
A-5
50 kHz–
100 kHz–
100 kHz
300 kHz
0.6 + 0.08
4.0 + 0.5
0.6 + 0.08
4.0 + 0.5
0.01 + .006
0.03 + .01

Advertisement

Table of Contents
loading

Table of Contents