Table 1-13
Limits for plug-in module resistance verification
Nominal resis-
Ω Range
tance
100Ω∗
100Ω
1kΩ
1kΩ
10kΩ
10kΩ
100kΩ
100kΩ
1MΩ
1MΩ
10MΩ
10MΩ
100MΩ
100MΩ
* Enable OCOMP for 100Ω range.
** Calculate limits based on actual calibration resistance values and Model 2700 one-year resistance accuracy specifications. See
Verification limits.

Verifying temperature

Thermocouple, thermistor, and RTD temperature readings are derived from DC volts and
resistance measurements respectively. For that reason, it is not necessary to independently ver-
ify the accuracy of temperature measurements. As long as the DC volts and resistance functions
meet or exceed specifications, temperature function accuracy is automatically verified. How-
ever, temperature verification procedures are provided below for those who wish to separately
verify temperature accuracy.
Thermocouple temperature
1.
2.
3.
4.
Nominal reading limits
(1 year, 18°C to 28°C)
99.9884 to 100.0116Ω
0.999894 to 1.000106kΩ
9.99894 to 10.00106kΩ
99.9890 to 100.0110kΩ
0.999890 to 1.000110MΩ
9.99590 to 10.00410MΩ
99.5770 to 100.4230MΩ
Connect the DC voltage calibrator output terminals and ice point reference to the
Model 7700 CH1 H and L INPUT terminals using low-thermal shielded connections, as
shown in
Figure
1-14.
Install the Model 7700 in Slot 1 of the Model 2700, then turn on the power, and allow
the unit to warm up for two hours before proceeding. Be sure the front panel INPUTS
switch is set to the REAR position.
Select the temperature function by pressing the TEMP key. Close Channel 1 by press-
ing the CLOSE key and keying in 101.
Configure the Model 2700 for °C units, type K temperature sensor, and internal refer-
ence junction as follows:
a.
Press SHIFT then SENSOR, and note the unit displays the temperature units:
UNITS: C. (If necessary, use the cursor and range keys to select °C units.)
b.
Press ENTER. The unit then displays the sensor type: SENS: TCOUPLE.
c.
Make sure that TCOUPLE is displayed, then press ENTER. The unit displays the
thermocouple type: TYPE: J.
d.
Select a type K temperature sensor, then press ENTER. The unit then displays the
reference junction type: JUNC: SIM.
e.
Select INT reference junction, then press ENTER.
Performance Verification
Recalculated limits**
__________ to __________ Ω
__________ to __________ kΩ
__________ to __________ kΩ
__________ to __________ kΩ
__________ to __________ MΩ
__________ to __________ MΩ
__________ to __________ MΩ
1-27