Application of the ankle cuff

(in ABI mode)
1) The patient has to lie down in supine position.
2) Place the ankle cuff on the leg. Make sure the
edge of the ankle cuff lies approximately 2 to 3cm
( ¾ to 1 inch) above the ankle and notice that the
artery mark is on the posterior tibial artery.
posterior tibial artery.
9
3) Wrap and tighten the cuff around the leg.
4) Leave a little free space between the leg of
the patient and the cuff. Two fingers should fit
between the leg and the cuff.
2-3 cm
(3/4 to 1 inch)

Three operation modes

ROUTINE
Select an operation mode
AUSCULT.
SCREEN
The WatchBP Office ABI device includes three
IAD
MAP
operation modes: «ABI», «ROUTINE», and «SCREEN»
(Inter-arm difference). Use the Mode Switch to select
the desired mode.
ABI ROUTINE SCREEN
ABI ROUTINE SCREEN
ABI ROUTINE SCREEN
«ABI» Mode
Select «ABI» Mode for Ankle-Brachial
measurement. Select the arm with the higher
blood pressure value as determined with the
«SCREEN» Mode
ABI ROUTINE SCREEN
ABI ROUTINE SCREEN
ABI ROU
10