Taking measurements using WatchBP Office ABI (cont.)
5) Press the M Button to switch between left(L) and
right(R). Press the Start/Stop Button to save the
setting.
ROUTINE
ROUTINE
6) Start measurements – Press the Start/Stop
Button to start the measurements.
ROUTINE
15-second
countdown...
17
7) Three consecutive measurements –
The device will take three consecutive
measurements with 15 second intervals in
between on default.
2
1
15
15
seconds
seconds
8) Inflation by fuzzy logic – The device will
adjust inflation pressure by fuzzy logic
technology.
9) The average value of the measurements–
The average values will be displayed when
the measurements are completed and will be
saved to the memory automatically.
The user can manually set the measurement intervals
*
of 15, 30, 45 or 60 seconds in ROUTINE Mode. (Please
refer to special functions section page 19 "Setting up
measurement intervals").
Skipping the countdown time
The 60 seconds countdown before measurement in
«SCREEN» Mode and 15 seconds in «ROUTINE» Mode
can be skipped by pushing the Start/Stop Button.
While the Start/Stop Button is pushed, the device will
3
immediately begin the next measurement.
Skip the countdown time
and begin measurement.
The device can be set into Stand-by Mode by pressing
*
the Start/Stop Button after the completion of
measurements. The device will automatically switch to
Stand-by Mode if left unattended for five minutes.
Stand by mode
Taking less than three measurements
The measurement sequence can be stopped
at anytime by pushing the Start/Stop Button
«SCREEN» Mode or «ROUTINE» Mode. The device
enters stand-by and the remaining measurements
are cancelled. Average of the measurements is
saved and can be viewed by pushing the M Button.
Cancel remaining
measurements at
anytime during
the measurement
sequence.