6) Start measurements – Press the Start/Stop
Button to start the measurement.
ABI
7) Inflation by fuzzy logic – The device will
adjust inflation pressure automatically by fuzzy
logic technology.
21
8) Display of the measurement values–
The measurement values will be displayed
and the Ankle-brachial index is automatically
calculated when the measurement is completed.
The Ankle-brachial index and the blood pressure
value display first, the brachial index value is
replaced by the pulse value for 3 seconds.
All the values are saved automatically.
ABI
ABI
INDEX

Special Functions

Hiding measurement values

The WatchBP Office ABI device features a Hide
function in order to prevent unnecessary elevated
blood pressure in patients due to nervousness
triggered by visible blood pressure values.
This feature is only available in ROUTINE mode
*
When the «Hide» function is ON, the countdown
*
between each measurement and the Hide icon will
display.
1) Activate the «Hide» Function – Press M button;
L or R flashes; keep pressing and holding the M
Button for 7 more seconds until On or Off flashes.
Press and hold
L or R flashes...
for 7 seconds...
2) Select ON or OFF – Press the M Button again to
turn the "Hide" function ON or OFF.
3) Confirm – Press Start/Stop to confirm the setting.
The default setting of "Hide" in «SCREEN» Mode is set
*
to "ON" and cannot be set to "OFF".