EN
The Microlife WatchBP Home A is designed to provide reliable and
Read the instructions carefully before using this device.
unbiased blood pressure measurement results, and strictly follows
European Society of Hypertension (ESH) and American Heart Asso-
ciation (AHA) recommendations for out-of-office blood pressure
measurement.
1
The WatchBP Home A device has been clinically validated according
to the ESH protocol.
O'Brien E, Asmar R, Beilin L, Imai Y, et al. European Society of
1
Hypertension recommendations for conventional, ambulatory and
home blood pressure measurement. European Society of Hyperten-
sion Working Group on Blood Pressure Monitoring. J Hypertens
2003;21:821-848.
WatchBP Home A
1
EN

Advertisement

Table of Contents
loading

Summary of Contents for Microlife WatchBP Home A

  • Page 1 Microlife WatchBP Home A The Microlife WatchBP Home A is designed to provide reliable and Read the instructions carefully before using this device. unbiased blood pressure measurement results, and strictly follows European Society of Hypertension (ESH) and American Heart Asso- ciation (AHA) recommendations for out-of-office blood pressure measurement.
  • Page 2: Table Of Contents

    • Deleting measurements 1. Product description • Name of parts and display • Transferring measurements 2. Before using WatchBP Home A for the first time 9. Battery indicator and battery change • Activating the device • Battery indicator • Setting the date and time 10.
  • Page 3: Product Description

    1. Product description Name of parts and display USUAL DIAG. NOCTURNAL 1 Cuff socket 5 M button (Memory) 2 Serial port 6 Display 3 Mode switch 7 ON/OFF button 4 Power socket 8 Battery compartment WatchBP Home A...
  • Page 4 Display 9 Mode indication AT Doctor symbol AK Atrial fibrillation detection (Afib) AL Outside measurement time AM Relax AN Battery display AO Delete memory data AP Stored value AQ Morning data AR Evening data AS Date/Time BT Systolic value BK Diastolic value BL Pulse indicator BM Pulse rate BN Number of stored data...
  • Page 5: Before Using Watchbp Home A For The First Time

    2. Before using WatchBP Home A for the first time Activating the device Pull out the protective strip from the battery compartment. 2. Set the month: Use the M button to set the month. Press the ON/ OFF button to confirm.
  • Page 6: Selecting The Correct Cuff

    Complete the process as described in previous steps. Selecting the correct cuff XL (X-Large size) The WatchBP Home A device is available with different cuff sizes. If the cuff provided with the device is an unsuitable size, please contact 32 - 52 cm (12.6 - 20.5 inches) Microlife.
  • Page 7: Selecting The Operation Mode

    Prior to each measurement, use the Mode switch on the side of the device to select the proper measurement mode. Choose out of 2 measurement modes: «DIAG.» (Diagnostic) or «USUAL» (Usual) mode. DIAG. USUAL WatchBP Home A...
  • Page 8: Diag.» Mode

     Two sets of measurements per day: ESH guidelines recom- «DIAG.» mode  Recommended by doctors: Blood pressure is measured in mend one double measurement taken in the morning between accordance with the measurement guidelines of the European 06:00 - 09:00 and one in the evening between 18:00 - 21:00. Society of Hypertension (ESH).
  • Page 9  Extended measurement period: WatchBP Home A has an  Evaluation: After measurements have been carried out for a total extended measurement period and allows morning measurements of 7 working days, take the device to your doctor for evaluation of between 04:00 - 12:00 and evening measurements between 18:00 your home blood pressure data.
  • Page 10: Usual» Mode

    4. Taking measurements using WatchBP Home A «USUAL» mode  With Afib detection: In «USUAL» mode, three consecutive Easy steps for measuring blood pressure properly measurements are taken automatically at 15 second intervals. 1. Avoid taking measurements directly after eating, drinking or smoking.
  • Page 11 2-3 cm from the inner fold of your arm. The tube connecting the cuff to the device should be placed on the inside of the arm. (Additional visual instruction can be found on the cuff). WatchBP Home A...
  • Page 12 4. Sit down and relax for at least 5 minutes prior to the measurement. 5. Sit upright and lean comfortably against the chair’s backrest. Press the ON/OFF button. The device will initiate a 60-second countdown in «DIAG.» mode and has no countdown in «USUAL» mode. ...
  • Page 13 The second measurement will start after 60-seconds. During this time avoid any movement. Automatically stored 60 sec.  If an error displays after the readings, please repeat the first 6 steps once again. WatchBP Home A...
  • Page 14: Data Memory

    8. When seven days of measurements have been collected, the doctor symbol will flash on the display. The WatchBP Home A can store up to 250 averaged measurement readings in «USUAL» mode. In «DIAG.» mode the memory is full after the schedule has been completed.
  • Page 15: Atrial Fibrillation Detection (Afib)

    * Joseph Wiesel, et al. Detection of Atrial Fibrillation Using a Modified Atrial fibrillation is a rhythm problem that can last from a few minutes, Microlife Blood Pressure Monitor. American Journal of Hypertension to days or weeks and even years. Atrial fibrillation can lead to the 2009;...
  • Page 16: Frequent Screening On Multiple Days

    Frequent screening on multiple days Afib icon is displayed  If atrial fibrillation is detected during all readings of the triple Use this device regularly, once per week, or once per month to screen for atrial fibrillation. measurements in «USUAL» mode, at both readings of two Some people may have atrial fibrillation occasionally that lasts longer measurements or all four readings of one day in «DIAG.»...
  • Page 17: Information For The Doctor

    2. Then press the M Button. click on «setup.exe» in the CD’s directory. 2. Follow the instructions provided in the installation window on the computer screen. 3. When the installation is finished, be sure to restart the computer before using the program. WatchBP Home A...
  • Page 18 In «DIAG.» mode 2. Press the M button again to display the average of all morning 1. When the M button is pressed, it briefly displays the total number data. Press the M button once again to show the average of all of measurements stored, e.g.
  • Page 19: Deleting Measurements

    3. All individual readings can be viewed by repeatedly pressing the M button. Ask your doctor if the measurement data is still required, before deleting data. Only delete the stored measurements when you are sure that you no longer need the data. Individual reading WatchBP Home A...
  • Page 20 1. Use the Mode switch to select the mode of measurements you 3. Release the M button and press it once more while the delete want to delete. symbol flashes. The deleting is confirmed by the beep sound. DIAG. USUAL 2.
  • Page 21: Transferring Measurements

    System Requirements: 550MHz CPU. 256MB Memory, 1024x768 pixel resolution, 256 color, CD-ROM drive, 1 free USB port, 40MB free hard disk space, Microsoft Windows XP / Vista / Win 7.  Or download the latest version from: www.watchbp.com. WatchBP Home A...
  • Page 22 Software commands Refer to the software user manual for detailed information and instructions Transfer «DIAG.» mode data Click «Download» Corresponding value You can remove the check mark and the corresponding value will not be used to calculate the average. Store data Click «Save», the file name is formed automatically from the patient’s identity number and the suffix «Dmode.xls».
  • Page 23: Battery Indicator And Battery Change

     Use 4 new, long-life 1.5V, size AA batteries.  Do not use batteries beyond their date of expiry.  Remove batteries, if the device will not be used for a prolonged period. WatchBP Home A...
  • Page 24: Safety, Care, Accuracy Test And Disposal

    10. Safety, care, accuracy test and disposal Using a power adaptor The WatchBP Home A can also be operated using a Microlife power  This device may be used only for the purpose described in this adaptor (DC 6V, 600 mA).
  • Page 25: Device Care

    Never open device. Protect device from impact and drops. Accuracy test We recommend the WatchBP Home A device be tested for accuracy every 2 years or after mechanical impact (e.g. being dropped). Please contact Microlife to arrange for an accuracy test.
  • Page 26: Cleaning The Cuff

    11. Error messages Cleaning the cuff Carefully remove spots on the cuff with a damp cloth and soapsuds. If an error occurs during measurement, the measurement is inter- rupted and an error message «ERR» is displayed. Do not wash the cuff! Do not iron the cuff! Disposal Batteries and electronic devices must be disposed of in accor-...
  • Page 27 Repeat the muscle tension. Repeat the measure- measurement.* ment, keeping your arm still. * Please contact Microlife, if this or any other problem occurs repeat- «ERR 3» No pressure in An adequate pressure cannot be edly. the cuff generated in the cuff.
  • Page 28: Important Facts About Blood Pressure And Home Measurements

    12. Important facts about blood pressure and home  Always discuss your home blood pressure measurement data with your doctor and tell him/her if you have noticed anything measurements unusual or feel unsure. Never rely on single blood pressure readings. ...
  • Page 29: Technical Specifications

    Static accuracy: pressure within ± 3 mmHg a readout value between 150/85 or 120/98 mmHg indicates «Grade Pulse accuracy: ± 5 % of the readout value 1 hypertension». Voltage source: 4x 1.5 V batteries; size AA Mains adapter DC 6V, 600 mA (optional) WatchBP Home A...
  • Page 30: Guarantee Card

    Product: WatchBP Home A EEC for Medical. Devices Class IIa have Product Number: BP 3MX1-3 been fulfilled. Serial Number: Date: Microlife reserves the right to alter technical specifications without prior written notice. Name: Address: Type BF applied part Reference number...
  • Page 31 Taipei, 11492, Taiwan, R.O.C. Clearwater, FL 33755, USA Tel. +886 2 8797 1288 Tel. +1 727 442 5353 Fax +886 2 8797 1283 Fax +1 727 442 5377 Email: watchbp@microlife.com.tw Email: msa@microlifeusa.com www.watchbp.com www.watchbp.com IB WatchBP Home A EN 3113 WatchBP Home A...

Table of Contents