Device care
Clean the device with a soft, dry cloth.
Accuracy test
We recommend the WatchBP Office ABI device
be tested for accuracy every 2 years or after
mechanical impact (e.g. being dropped). Please
contact Microlife to arrange for an accuracy test.
37
Cleaning the cuff
DO NOT wash the cuff. DO NOT iron the cuff cover.
Do not wash the cuff!
Do not iron the cuff!
Disposal
Batteries and electronic instruments must be
disposed of in accordance with the locally applicable
regulations, and not as domestic waste.

Technical specifications

Operation temperature/
humidity:
10 to 40 °C (50 to 104 °F)
Storage temperature/
-20 to 55 °C (-4 to 131 °F)
humidity:
15 - 90 % relative maximum humidity
Weight:
1100 g (including rechargeable
battery pack)
Dimensions:
200 x 125 x 90 mm
Measuring method:
Oscillometric, corresponding to
Korotkoff
Measurement range:
30 - 280 mmHg – blood pressure
40 - 200 beats per minute – pulse
Cuff pressure display:
Range: 0 - 299 mmHg
Resolution: 1 mmHg
Static accuracy: pressure within ± 3
mmHg or 2% of reading > 200mmHg
Pulse accuracy: ±5 % of the readout
value
Voltage source:
Rechargeable battery pack;
4.8V C4000 mAh
Mains adapter DC 7.5V, 2 A
Reference to Standards: • Device corresponds to the
requirements of the standard for non-
invasive blood pressure monitor.
EN 1060-1
EN 1060-3
EN 1060-4
IEC 60601-1
IEC 60601-1-2
Electromagnetic
• Device fulfills the stipulations of the
standard IEC 60601-1-2.
compatibility:
The stipulations of the EU Directive 93/42/
EEC for Medical Devices Class IIa have been
fulfilled.
Type BF applied
Reference number
part
Serial number
Manufacturer
Microlife reserves the right to alter technical
specfications without prior written notice.
38