Important Control Solution Test Information

˙ Use only FORA control solutions.
˙ Check the expiration date on the control solution vial. Do not use if expired.
˙ Control solution, meter, and test strips should come to room temperature (68
to 77°F/20 to 25°C) before testing.
˙ Shake the vial, discard the first drop of control solution, and wipe off the dis-
penser tip to ensure a clean sample and an accurate result.
˙ Use only 90 days after first opening. Record the discard date (date opened
plus 90 days) on the control solution vial. Discard after 90 days.
˙ Store the control solution tightly closed at temperatures 36°F to 86°F (2°C to
30°C). Do not freeze.
The control solution range printed on the test strip vial is for control
solution only. It is used to test meter and test strip performance. It is not a
recommended range for your blood glucose level.