Important Control Solution Information

˙ Use only FORA control solutions.
˙ Check the expiration date on the control solution vial. Do not use if expired.
˙ Control solution, meter, and test strips should come to room temperature 68 to
77°F(20 to 25°C) before testing.
˙ Shake the vial, discard the first drop of control solution, and wipe off the dis-
penser tip to ensure a good sample and an accurate result.
˙ Use only for 90 days after first opening. Record the discard date (date opened
plus 90 days) on the control solution vial. Discard after 90 days.
˙ Store the control solution tightly closed at temperatures 36 to 86°F (2 to 30°C).
Do not freeze.
The control solution range is printed on FORA test strip vial. Only use
FORA control solution. It is used to test meter and test strip performance.
It is not recommended range for your blood glucose level.
28