Important Control Solution Test Information

►Use only FORA control solutions. Check the expiration date on the
control solution vial. Do not use if expired.
►Control solution, meter, and test strips should be at room
temperature, 68-77°F (20-25°C), before testing.
►Shake the vial before use. Discard the first drop of control solution.
After squeezing, wipe off the dispenser tip to avoid contamination.
The above tips ensure you to get a clean sample and an accurate
result.
►Use only 3 months after the first opening date. Record the discard
date (date opened plus 90 days) on the control solution vial. Dis-
card the vial after 90 days.
►Store the control solution with lid tightly closed at temperatures
36-86°F (2-30°C). Do not freeze.
The control solution range is printed on FORA test strip vial. Only use
FORA control solution with FORA test strips. It is used to evaluate the
meter and test strip performance. It is not a recommended range for your
blood glucose level.