Adjustment With External Weight; Models Cpb (Non Verifiable Models) - KERN CPB 3K0.1 Operating Manual

Hide thumbs Also See for CPB 3K0.1:
Table of Contents

Advertisement

7.7 Adjustment with external weight

As the acceleration value due to gravity is not the same at every location on earth,
each balance must be coordinated - in compliance with the underlying physical
weighing principle - to the existing acceleration due to gravity at its place of location
(only if the balance has not already been adjusted to the location in the factory). This
adjustment process must be carried out for the first commissioning, after each
change of location as well as in case of fluctuating environment temperature. To
receive accurate measuring values it is also recommended to adjust the balance
periodically in weighing operation.
Procedure when adjusting:
Observe stable environmental conditions. A warming up time of ca. 2 h is required for
stabilization. Ensure that there are no objects on the weighing plate.

7.7.1 Models CPB (Non verifiable models)

Operation
Switch on balance and during self-check
press the TARE key
Use the number keys to enter password:
Either
• Default password "0000"
or
• Personal password, enter under
[
function
F6 PKn
Ensure that there are no objects on the
weighing plate.
50
]
Confirm
Display
PKN
NET
a
Piece weight
----
PRESET
F1 CAL
NET
a
VnLQAD
CPB-BA-defsi-0712
Weight
Weight

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cpb 30 k1

Table of Contents