7.9.2 External adjustment with user-defined adjustment weight < caleud >
20
 Place the adjustment weight and confirm with ➔ button,
< wait> followed by < reMvld> will be displayed.
 Once < reMvld> is displayed, remove the adjustment
weight.
 After successful adjustment the balance automatically re-
turns to weighing mode.
In case of an adjustment error (e.g. objects on the weighing
plate) the display will show the error message < wrong
>. Switch off balance and repeat the adjustment process.
 Press and hold the TARE and ON/OFF buttons simultane-
ously to enter the setup menu.
 Wait until the first menu item < Cal > is displayed.
 Confirm by ➔ button, < CalExt> will be displayed.
 Use the navigation keys to select   < caleud>.
 Acknowledge by ➔ button. The numeric input window for
the weight value of the adjustment weight appears. The ac-
tive digit is flashing.
 Provide adjustment weight.
 Enter weight value, numerical input see chap. 3.2.2