Models Cpb - KERN CPB 3K0.1 Operating Manual

Hide thumbs Also See for CPB 3K0.1:
Table of Contents

Advertisement

Zero point will be saved.
Confirm weighing value
Wait until the stability display shows:
Adjustment process is started.
The process of adjustment is completed.
Remove adjusting weight, balance will return
into weighing mode automatically. In case of
an adjustment error or incorrect adjusting
weight the display will show an error
message (SPAN / FAKLD); repeat adjustment
process.
7.7.2 Models CPB.. M (verifiable models)
For verified balances adjustment by means of a jumper is locked. In order to carry
out adjustment the two contacts of the conductor plate must be shorted with a jumper
(see chapt. 7.8.1).
Observe stable environmental conditions. A warming up time of ca. 2 h is required for
stabilization. Ensure that there are no objects on the weighing plate.
CPB-BA-defsi-0712
SEL
15
The balance will suggest a
weighing value for adjusting
weight. However, you may enter a
value of your choice via the
number keys.
LQAD 15
Carefully place adjusting weight in
the centre of the weighing plate
SPAnPASS
000000
999999
kg
kg
0.0
51

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cpb 30 k1

Table of Contents