Cpb-Ba-Defsi - KERN CPB 3K0.1 Operating Manual

Hide thumbs Also See for CPB 3K0.1:
Table of Contents

Advertisement

Error message
Zeroing only in the permitted
Err 4
range (4% of the max.
weighing range) and not for
overload/underload
Keyboard error
Err 5
Value outside the A/D changer
Err 6
range
Adjustment error
FAIL H or
FAIL L
Adjustment error
Err 8
Err 9
Weighing result unstable
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.

CPB-BA-defsi-0712

Description
Possible causes
• Object on the weighing
plate
• Overload when zeroing
• Improper adjustment
• Damaged weighing cell
• Damaged electronics
• Improper operation of the
balance
• Weighing plate not
installed
• Damaged weighing cell
• Damaged electronics
• Improper adjustment
• Deviation to factory
adjustment > 10%
• Improper adjustment
• Incorrect adjusting weight
• Balance unstable
• Draught/air movement
• Table/floor vibrations
71