Errors are retrieved in first-in-first-out ( FIFO ) order. The first error
returned is the first error that was stored. Errors are cleared as you
read them. When you have read all errors from the queue, the ERROR
annunciator turns off and the errors are cleared. The instrument
beeps once each time an error is generated.
If more than 10 errors have occurred, the last error stored in the
queue (the most recent error) is replaced with "Error queue overflow".
No additional errors are stored until you remove errors from the
queue. If no errors have occurred when you read the error queue,
the instrument responds with "No error".
The error queue is cleared by the *CLS (clear status) command or when
power is cycled. The errors are also cleared when you read the queue.
The error queue is not cleared by a Factory Reset (*RST command)
or an Instrument Preset (SYSTem:PRESet command).
Front-Panel Operation:
ERRORS
If the ERROR annunciator is on, press
knob to scroll through the error numbers. Press
the error message. Press
(the final key press cancels the scroll). All errors are cleared when
you exit the menu.
Remote Interface Operation:
SYSTem:ERRor?
Errors have the following format (the error string may contain
up to 80 characters):
-113,"Undefined header"
304
to view the errors. Use the
to view the text of
again to increase the scrolling speed
Read and clear one error from the queue