Chapter 6 Error Messages

Execution Errors

-168
Block data not allowed
Data was sent to the instrument in SCPI definite length block format but
this command does not accept this format. Example: SOUR:DIG:DATA #128
-178
Expression data not allowed
A channel list was received but is not allowed for this command.
Example: SYST:CTYPE? (
-211
Trigger ignored
More than one trigger was received while the instrument was scanning.
Triggers are occurring too frequently and you may need to slow them
down. Also make sure that you have selected the proper trigger source.
INIT ignored
-213
An INITiate command was received but could not be executed because
a scan was already in progress. Send an ABORt command or bus Device
Clear to stop a scan in progress.
Trigger deadlock
-214
A trigger deadlock occurs when the trigger source is "BUS" and a READ?
command is received.
Settings conflict
-221
An invalid configuration was requested. This error is most commonly
generated when setting alarm limits. Note that the lower limit must
always be less than or equal to the upper limit, even if you are using
only one of the limits. This error is also generated if you send the
MEASure? or CONFigure command with autorange enabled with a
fixed resolution.
-222
Data out of range
A numeric parameter value is outside the valid range for this command.
Example: TRIG:COUNT -3
-223
Too much data
A character string was received but could not be executed because the
string length was more than 12 characters. This error can be generated
by the CAL:STRing and DISPlay:TEXT commands.
-224
Illegal parameter value
A discrete parameter was received which was not a valid choice for this
command. You may have used an invalid parameter choice.
Example: TRIG:SOURCE ALARM
@
100)
(ALARM is not a valid choice)
6
307