Chapter 9 Specifications; Dc, Resistance, And Temperature Accuracy Specifications - HP 34970A User Manual

Data acquisition / switch unit
Table of Contents

Advertisement

Chapter 9 Specifications

DC, Resistance, and Temperature Accuracy Specifications

DC, Resistance, and Temperature Accuracy Specifications
[1]
( % of reading + % of range )
Includes measurement error, switching error, and transducer conversion error
[3]
Function
Range
DC Voltage
100.0000 mV
1.000000 V
10.00000 V
100.0000 V
300.000 V
[4]
Resistance
100.0000
1.000000 k
10.00000 k
100.0000 k
1.000000 M
10.00000 M
100.0000 M
DC Current
10.00000 mA
100.0000 mA
HP 34901A Only
1.000000 A
Temperature
Type
Thermocouple
B
E
J
K
N
R
S
T
RTD
R
from 49
0
to 2.1 k
Thermistor
2.2 k, 5 k, 10 k
[1] Specifications are for 1 hour warm up and 6
[2] Relative to calibration standards
[3] 20% over range on all ranges except 300 Vdc and 1 Adc ranges
[4] Specifications are for 4-wire ohms function or 2-wire ohms using Scaling to remove the offset.
Without Scaling, add 1 additional error in 2-wire ohms function.
[5] 1 year accuracy. For total measurement accuracy, add temperature probe error.
404
Test Current or
24 Hour
±
Burden Voltage
23 °C
0.0030 + 0.0035
0.0020 + 0.0006
0.0015 + 0.0004
0.0020 + 0.0006
0.0020 + 0.0020
1 mA current source
0.0030 + 0.0035
1 mA
0.0020 + 0.0006
100 A
0.0020 + 0.0005
0.0020 + 0.0005
10 A
5 A
0.002 + 0.001
500 nA
0.015 + 0.001
500 nA || 10 M
0.300 + 0.010
< 0.1 V burden
0.005 + 0.010
< 0.6 V
0.010 + 0.004
< 2 V
0.050 + 0.006
[5]
Best Range Accuracy
1100 C to 1820 C
1.2 C
-150 C to 1000 C
1.0 C
-150 C to 1200 C
1.0 C
-100 C to 600 C
1.0 C
-100 C to 1300 C
1.0 C
300 C to 1760 C
1.2 C
400 C to 1760 C
1.2 C
-100 C to 400 C
1.0 C
-200 C to 600 C
0.06 C
-80 C to 150 C
0.08 C
1
digits
2
[2]
90 Day
±
1 °C
23 °C
5 °C
23 °C
0.0040 + 0.0040
0.0050 + 0.0040
0.0030 + 0.0007
0.0040 + 0.0007
0.0020 + 0.0005
0.0035 + 0.0005
0.0035 + 0.0006
0.0045 + 0.0006
0.0035 + 0.0030
0.0045 + 0.0030
0.008 + 0.004
0.010 + 0.004
0.008 + 0.001
0.010 + 0.001
0.008 + 0.001
0.010 + 0.001
0.008 + 0.001
0.010 + 0.001
0.008 + 0.001
0.010 + 0.001
0.020 + 0.001
0.040 + 0.001
0.800 + 0.010
0.800 + 0.010
0.030 + 0.020
0.050 + 0.020
0.030 + 0.005
0.050 + 0.005
0.080 + 0.010
0.100 + 0.010
Extended Range Accuracy
400 C to 1100 C
-260 C to -150 C
-210 C to -150 C
-230 C to -100 C
-220 C to -100 C
-50 C to 300 C
-50 C to 400 C
-240 C to -100 C
Temperature
1 Year
Coefficient /°C
±
5 °C
0 °C – 18 °C
28 °C – 55 °C
0.0005 + 0.0005
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0001
0.0005 + 0.0003
0.0006 + 0.0005
0.0006 + 0.0001
0.0006 + 0.0001
0.0006 + 0.0001
0.0010 + 0.0002
0.0030 + 0.0004
0.1500 + 0.0002
0.002 + 0.0020
0.002 + 0.0005
0.005 + 0.0010
[5]
1.8 C
0.03 C
1.5 C
0.03 C
1.2 C
0.03 C
1.5 C
0.03 C
1.5 C
0.03 C
1.8 C
0.03 C
1.8 C
0.03 C
1.5 C
0.03 C
0.003 C
0.002 C

Advertisement

Table of Contents
loading

Table of Contents