HP 34970A User Manual page 110

Data acquisition / switch unit
Table of Contents

Advertisement

Chapter 4 Features and Functions
Temperature Measurement Configuration
Remote Interface Operation: You can use the MEASure? or
CONFigure command to select the probe type and RTD type.
For example, the following statement configures channel 301 for
2-wire measurements of an RTD with
= 0.00385 or "91" to specify
CONF:TEMP RTD,85,(
You can also use the SENSe command to select the probe type,
type, and nominal resistance. For example, the following
RTD
statement configures channel 103 for 4-wire measurements of an
with
= 0.00391 (channel 103 is automatically paired with
RTD
channel 113 for the 4-wire measurement).
SENS:TEMP:TRAN:FRTD:TYPE 91,(
The following statement sets the nominal resistance (R
on channel 103.
SENS:TEMP:TRAN:FRTD:RES 1000,(
= 0.00385 (use "85" to specify
= 0.00391).
@
301)
@
103)
@
103)
4
) to 1000
0
111

Advertisement

Table of Contents
loading

Table of Contents