Scanning
Remote Interface Operation: The following program segment configures
the instrument to scan when an alarm occurs.
TRIG:SOURCE ALARM1
TRIG:COUNT 2
CALC:LIM:UPPER 5,(
CALC:LIM:UPPER:STATE ON,(
OUTPUT:ALARM1:SOURCE (
ROUT:MON (
@
103)
ROUT:MON:STATE ON
INIT
Note: To stop a scan, send the ABORt command.
Select the alarm configuration
Sweep the scan list 2 times
Set the upper limit
@
103)
103) Enable the upper limit
@
Report alarms on Alarm 1
@
103)
Select monitor channel
Enable monitoring
Initiate the scan