Chapter 6 Error Messages

Instrument Errors

Instrument Errors
Channel list: slot number out of range
111
The specified slot number is invalid. The channel number has the
form (
channel number. Example: CONF:VOLT:DC (
112
Channel list: channel number out of range
The specified channel number is invalid for the module in the selected slot.
The channel number has the form (
(100, 200, or 300) and cc is the channel number.
Example: ROUT:CLOSE (
113
Channel list: empty scan list
Before you can initiate a scan, you must set up a scan list which includes
all configured multiplexer or digital channels in the instrument. Use the
MEASure?, CONFigure, or ROUTe:SCAN commands to set up your scan list.
201
Memory lost: stored state
This error is reported at power-on to indicate that a stored state has
become unusable. This error is most likely caused by a dead battery
(memory is battery-backed). Refer to the HP 34970A Service Guide to
replace the internal battery.
Memory lost: power-on state
202
This error is reported at power-on to indicate that the power-down state
of the instrument (normally recalled when power is turned on) has become
unusable. This error is most likely caused by a dead battery (memory is
battery-backed). Refer to the HP 34970A Service Guide to replace the
internal battery.
203
Memory lost: stored readings
This error is reported at power-on to indicate that readings stored in
memory from a previous scan have been lost. This error is most likely
caused by a dead battery (memory is battery-backed). Refer to the
HP 34970A Service Guide to replace the internal battery.
204
Memory lost: time and date
This error is reported at power-on to indicate that the time and date
settings have been lost (they are reset to JAN 1, 1996 00:00:00 ).
This error is most likely caused by a dead battery (memory is
battery-backed). Refer to the HP 34970A Service Guide to replace the
internal battery.
scc), where s is the slot number (100, 200, or 300) and cc is the
@
@
134)
@
404)
scc), where s is the slot number
@
6
309

Advertisement

Table of Contents
loading

Table of Contents