To Store The Instrument State - HP 34970A User Manual

Data acquisition / switch unit
Table of Contents

Advertisement

Chapter 2 Front-Panel Overview

To Store the Instrument State

To Store the Instrument State

You can store the instrument state in one of five non-volatile storage
locations. A sixth storage location automatically holds the power-down
configuration of the instrument. When power is restored, the instrument
can automatically return to its state before power-down (a scan in
progress before power-down will also be resumed).
1 Select the storage location.
From the front panel, you have the ability to assign names (up to 12
characters) to each of the five stored states.
The storage locations are numbered 1 through 5. The power-down
state is automatically stored and can be recalled from the front panel
(the state is named LAST PWR DOWN ).
2 Store the instrument state.
The instrument stores all channel configurations, alarm values, scaling
values, scan interval setups, and advanced measurement configurations.
48

Advertisement

Table of Contents
loading