Immunity
IEC 60601
test
test level
Power
3 A/m
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to
80 MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the B
Coronis 5MP
64
Level
Electromagnetic environment –
guidance
Not
Power frequency magnetic fields should be
applic
at levels characteristic of a typical location
able
in a typical commercial or hospital
environment.
3 V
Portable and mobile RF communications
equipment should be used no closer to any
part of the B
ARCO
cables, than the recommended separation
distance calculated from the equation
3 V/
applicable to the frequency of the
m
transmitter.
Recommended separation distance
d 
1
2 .
P
80 MHz to 800 MHz
d 
1
2 .
P
800 MHz to 2.5 GHz
d 
2
3 .
P
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,
should be less than the
a
compliance level in each frequency range.
Interference may occur in the vicinity of
equipment marked with symbol:
C
Display, including
ORONIS
C
ARCO
ORONIS
b