P
OWER CONNECTION
SUPPLY
Power requirements: The display must be powered using the
delivered medical approved 24 VDC SELV power supply.
The medical approved DC power supply must be powered by the
AC mains voltage (protective earth terminal).
T
-
RANSIENT OVER
VOLTAGE
If the equipment is not used for a long time, disconnect it from the AC
inlet to avoid damage by transient over-voltage.
To fully disengage the power to the unit, please disconnect the power
cord from the AC inlet.
P
:
OWER CORDS
Utilize a UL-listed detachable power cord, 3-wire, type SJ or
equivalent, 18 AWG min., rated 300 V min., provided with a
hospital-grade type plug 5-15P configuration for 120V application,
or 6-15P for 240V application.
Do not overload wall outlets and extension cords as this may result
in fire or electric shock.
Mains lead protection (U.S.: Power cord): Power cords should be
routed so that they are not likely to be walked upon or pinched by
items placed upon or against them, paying particular attention to
cords at plugs and receptacles.
W
ATER AND MOISTURE
Never expose the display to rain or moisture.
Never use the display near water - e.g. near a bathtub, washbasin,
swimming pool, kitchen sink, laundry tub or in a wet basement.
V
ENTILATION
Do not cover or block the ventilation openings in the cover of the set.
When installing the display in a cupboard or another closed location,
-
DISPLAY WITH EXTERNAL
24 VDC
POWER
Coronis 5MP
55