YOKOGAWA DpharpEJX vigilantplant EJX Series Manual

YOKOGAWA DpharpEJX vigilantplant EJX Series Manual

Hart communication type
Hide thumbs Also See for DpharpEJX vigilantplant EJX Series:
Table of Contents

Advertisement

Quick Links

User's
Manual
Yokogawa Electric Corporation
EJX Series
HART Communication Type
IM 01C25T01-01E
IM 01C25T01-01E
3rd Edition

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DpharpEJX vigilantplant EJX Series and is the answer not in the manual?

Questions and answers

Summary of Contents for YOKOGAWA DpharpEJX vigilantplant EJX Series

  • Page 1 User’s Manual EJX Series HART Communication Type IM 01C25T01-01E IM 01C25T01-01E 3rd Edition Yokogawa Electric Corporation...
  • Page 2: Table Of Contents

    3.5.8 Trim Analog Output ..............3-18 3.5.9 Burst Mode ................. 3-20 3.5.10 Multidrop Mode ................3-21 3.5.11 External Switch Mode ..............3-22 FD No. IM 01C25T01-01E IM 01C25T01-01E 3rd Edition: Oct. 2004 (YK) All Rights Reserved, Copyright © 2004, Yokogawa Electric Corporation...
  • Page 3 CONTENTS 3.5.12 CPU Failure Burnout Direction and Hardware Write Protect ..3-22 3.5.13 Software Write Protect ............... 3-23 (1) Setting Password ............... 3-23 (2) Entering Password to Enable the Parameter Changes ..... 3-23 (3) Releasing Password ..............3-23 (4) Software Seal ................3-24 3.5.14 Signal Characterizer ..............
  • Page 4: Introduction

    • All rights reserved. No part of this manual may be failure. reproduced in any form without Yokogawa’s written permission. • Yokogawa makes no warranty of any kind with regard to this manual, including, but not limited to, NOTE implied warranty of merchantability and fitness for a Draws attention to information essential for particular purpose.
  • Page 5: Safe Use Of This Product

    1. INTRODUCTION (e) Modification 1.1 Safe Use of This Product • Yokogawa will not be liable for malfunctions or For the safety of the operator and to protect the damage resulting from any modification made to this instrument and the system, please be sure to follow this instrument by the customer.
  • Page 6: Atex Documentation

    Se si desidera ricevere i manuali operativi di prodotti Ex in lingua locale, mettersi in Alle Betriebsanleitungen für ATEX Ex bezogene contatto con l’ufficio Yokogawa più vicino o con un Produkte stehen in den Sprachen Englisch, Deutsch rappresentante.
  • Page 7: Matching Of Communicator Dd And Instrument Dd

    1) Turn on only the communicator. 2) Call Utility from the main menu and press [→] . 3) Call Simulation and press [→] . 4) Select YOKOGAWA from the list of manufacturers by pressing [ ] ↓ and press [→] .
  • Page 8: Conditions Of Communication Line

    2. CONDITIONS OF COMMUNICATION LINE CONDITIONS OF COMMUNICATION LINE The HART communication signal is superimposed 2.2 Communication Line Re- onto the 4 to 20 mA DC analog signal. Since the quirements modulated wave is a communication signal, superim- posing it on the normal signal will, from basic prin- Specifications for communication line: ciples, cause no error in the DC component of the Supply voltage (general use type): 16.6 to 42 V DC...
  • Page 9: Operation

    3. OPERATION OPERATION 3.1 Basic Operation of the 275 HART Communicator 3.1.1 Keys and Functions Communication cable LCD (liquid crystal display) (21 characters×8 lines) Function keys Functions of the keys are indicated on the display. Pressing (HOME) when the display is as shown changes the display to the Online EJX: menu.
  • Page 10: Display

    Tag (8 characters) <a> current menu device, or mark because of a current menu data to send comunication error EJX :YOKOGAWA PGUP PGDN Online answer to move up one move down one answer to <b> 1 Device setup...
  • Page 11: Entering, Setting, And Sending Data

    4 LRV Display appears when the 5 URV ENTER HART Communicator is turned on. Select Device setup. Example: To change from Tag YOKOGAWA to FI1-1A. Call up the Tag setting display. EJX:YOKOGAWA Device setup ×2 1 Process Varlables 2 Diag/Service 1.
  • Page 12: Parameter Usage And Selection

    3. OPERATION 3.2 Parameter Usage and Selec- IMPORTANT tion After setting and sending data with the HART communicator, wait 30 seconds before turning Before setting a parameter, please see the following off the transmitter. If it is turned off too soon, the table for a summary of how and when each parameter settings will not be stored in the transmitter.
  • Page 13: Menu Tree

    3. OPERATION 3.3 Menu Tree (Diag/Service) (Device setup) (Process variables) 1 Test device 1 Status 1 Status group 1 1 Process variables 1 Pres 2 Self test 2 Status group 2 2 Pres % 3 Status group 3 3 AO 4 Status group 4 4 SP 5 Status group 5...
  • Page 14: Basic Setup

    ENTER HELP SAVE HOME F0309.EPS (3) Message Note that the Yokogawa default setting for the standard 1. Device setup EJX:YOKOGAWA temperature is 4 C (39.2 F). For the units of mmH Message O, and ftH O, the pressure varies according to the 3.
  • Page 15: Range Change

    Example: To change the range from 0 to 2500 mmH O to 500 to 3500 mmH Call up the Keypad input display. 1. Device setup 3. Basic setup 3. Re-range 1. Keypad input EJX:YOKOGAWA Keypad input 1 LRV 0.0 mmH2O 2 URV 2500.0 mmH2O 3 Unit mmH2O 4 LSL -10197.2 mmH2O...
  • Page 16: Apply Values

    1. Device setup cause the lower range value to change. Thus, chang- ing the upper range value also changes the span. 3. Basic setup 3. Re-range 2. Apply values EJX:YOKOGAWA WARN-Loop should be removed from automatic control (OK) Press OK (F4).
  • Page 17: Output Mode

    1. Device setup constant between 0.00 and 0.49 seconds. 3. Basic setup 5. Xfer fnctn Example: To change from 2.0 seconds to 0.5 seconds Call up the Pres Damp display. EJX:YOKOGAWA Xfer function Linear Linear 1. Device setup Sq root 3.
  • Page 18: Output Signal Low Cut Mode Setup

    Note that when the output modes of the output signal 9. H/L Swap and the display are selected as Sq root and Linear accordingly, the low cut function is not available for EJX:YOKOGAWA H/L Swap the display value. Normal Normal...
  • Page 19: Detailed Setup

    1. Device setup 1. Device setup 4. Detailed setup 4. Detailed setup 2. Signal condition 4. Display condition 9. Bi-dir mode 2. P disp condition EJX:YOKOGAWA 1. Disp Pres % fnctn Bi-dir mode EJX:YOKOGAWA Disp Pres % fnctn HELP SEND ENTER...
  • Page 20: Integral Indicator Scale Setup

    Indicates values of input pressure 1. Disp select (PRES) with the indication limits 99999 to 1. Disp Out 1 99999. PRES 456 kPa EJX:YOKOGAWA Disp Out 1 PRES % PRES % ENGR. PRES % of range Indicates input pressure in 2.5 to SP %...
  • Page 21 ‘– 5 0’ inHgA Nl/h g/cm3 EJX:YOKOGAWA Engr LRV FX0302.EPS (ENTER) ENTER Follow the procedure below to set your own unit. Call up the Engr LRV Display.Set –50, and press ENTER (F4). Example: Set the engineering unit as M/h.
  • Page 22: Unit For Displayed Temperature

    Example: Change the unit for the temperature display from degC to degF. 4. Detailed setup 1. Sensors 1. Device setup 2. SP sensor 4. Detailed setup 3. SP Unit 1. Sensors 3. Temp sensor EJX:YOKOGAWA SP Unit 3. Temp Unit mmH2O kg/cm2 torr EJX:YOKOGAWA HELP SEND ENTER...
  • Page 23: Test Output

    (-2.5%) to 21.6 mA (110%) . test output will continue for approximately 10 minutes. Example: To output 12 mA (50%) 1. Device setup 2. Diag/Service 2. Loop test EJX:YOKOGAWA WARN-loop should be removed from automatic control (OK) Set the control loop in manual ABORT mode, and press OK (F4).
  • Page 24: Sensor Trim

    Zeroing—Pres Zero trim 2. Pres Trim Pres Zero trim carries out the zero adjustment and automatically sets the applied “0” input values to the EJX:YOKOGAWA Select trim mode output value of “0,” keeping the span constant. Use 1 Off...
  • Page 25: Full Sensor Trim

    100% points within the 2. Pres Trim calibration range. In the manual method, the reference pressure should EJX:YOKOGAWA Select trim mode ×3 1 Off also be applied to the transmitter at both lower and...
  • Page 26: Sensor Trim For Static Pressure

    Example 1: For an adjustment using an ammeter (±1µA is measurable) 1. Device setup 2 Diag/Service 3. Calibration 2. Analog output trim 1. D/A trim EJX:YOKOGAWA Analog output trim 1 D/A trim 2 Scaled D/A trim 3 Clear D/A trim Select the D/A trim item. HELP...
  • Page 27 3. OPERATION Example 2: To adjust using a voltmeter EJX:YOKOGAWA Fld dev output 4.000 mA equal to reference 1. Device setup meter? (ENTER) 1 Yes 2 No 2 Diag/Service Ammeter reading: 4.000 HELP ABORT ENTER 3. Calibration If the reading on the ammeter is 2.
  • Page 28: Burst Mode

    3. OPERATION 3.5.9 Burst Mode ‘1 . 0 1’ EJX:YOKOGAWA When the burst mode is set on, the transmitter continu- Enter meter value 1.000000 ously sends stored data. Either the pressure value, % 1.010 (ENTER) range/current value, or current/process variables can be...
  • Page 29: Multidrop Mode

    3. OPERATION 3.5.10 Multidrop Mode Example: Communication when set in multidrop mode “Multidropping” transmitters refers to the connection of several transmitters to a single communications HART Communicator (1) The HART communicator transmission line. Up to 15 transmitters can be con- Online searches for a transmitter that is 1 EJX110A-1...
  • Page 30: External Switch Mode

    3. OPERATION 3.5.11 External Switch Mode 3.5.12 CPU Failure Burnout Direction and Hardware Write Protect Follow the procedure below to enable or inhibit zero point adjustment by means of the zero-adjustment There are two slide switches on the CPU assembly screw on the transmitter.
  • Page 31: Software Write Protect

    3. OPERATION 3.5.13 Software Write Protect EJX: EJX configured data is saved by using a write protect It changed the state of protection related function. The write protect status is set to “Yes” when password. (OK) 8 alphanumeric characters are entered in the New Press OK (F4).
  • Page 32: Software Seal

    Write Protect mode by 1 S.C. Disabled 2 Num of points (SEND) 3 point setting using the general use password: “YOKOGAWA.” 4 X Start 0.00 % 5 Y Start 0.00 % Press SEND (F2) to send the data...
  • Page 33: Process Alarm

    3. OPERATION 3.5.15 Process Alarm Example: Set the signal characterizer to Enabled. The function is used to display the alarm codes when 1. Device setup the input pressure exceeds the specified value within 4. Detailed setup the calibration range. The same is available for the 2.
  • Page 34: Status Output (Option Code Al)

    3. OPERATION 3.5.16 Status Output (option code AL) Example: Status output operation of ON WHEN AL. DETECT This feature is used for a transistor output (open Status output for higher alert value collector) of an on/off signal according to the status of Output high and low alarm limits, which are user-configurable 5% of hysteresis...
  • Page 35: Capillary Fill Fluid Density Compensation

    3. OPERATION 3.5.17 Capillary Fill Fluid Density Compensa- EJX118A tion For transmitters with diaphragm seals, this function is used to compensate the zero shift caused by the Transmitter ambient temperature effect on the capillary tubes. The following equation indicates the relationship between the calculated output value and the compen- sating constant K (%/ C) with the measured ambient EJX438A...
  • Page 36 3. OPERATION Table A. Constant value [B] of fill fluid Fill fluid A, C code 0.76 0.87 1.45 0.75 kgf/cm 0.000076 0.000087 0.000145 0.000075 0.00745 0.00853 0.01422 0.00736 mBar 0.07453 0.08532 0.14220 0.07355 0.000074 0.000084 0.000140 0.000073 0.02992 0.03425 0.05709 0.02953 0.00108 0.00124...
  • Page 37: Self-Diagnostics

    4. SELF-DIAGNOSTICS SELF-DIAGNOSTICS 4.1 Self-Diagnostics Diagnostic by “status” 1. Device setup 4.1.1 Identify Problems by Using the Communicator 2. Diag/Service 1. Test device The HART communicator can be used to run self- diagnostics on a transmitter and check for incorrect 1.
  • Page 38: Alarm Message Summary

    4. SELF-DIAGNOSTICS 4.2 Alarms and Countermeasures Table 4.2.1 Alarm Message Summary Integral HART communicator Output operation Status Cause Countermeasure indicator display during error group AL. 01 P sensor error Sensor problem. Outputs the signal (High or Low) Replace capsule when CAP.
  • Page 39 4. SELF-DIAGNOSTICS Table 4.2.2 HART Communicator Error Messages Error message Probable cause Countermeasure Change the setting. — Invalid selection Value was too high Set value is too high. Value was too low Set value is too low. — — Too few data bytes received —...
  • Page 40: Parameter Summary

    1, 3, 4, 2 Device information Dev id Device ID 1, 4, 5, 1, ↓, 6 Distributor Yokogawa 1, 4, 5, 1, ↓, 7 Drain vent matl Drain and vent plug material 1, 4, 5, 2, 5 Extra No. Customization number 1, 4, 5, 1, ↓, 4...
  • Page 41 5. PARAMETER SUMMARY Fast key Function Label Item Contents Default value Handling sequences Process variables Analog output current 3.6000 to 21.6000 mA 1, 1, 3 Engr Disp User scaled value 32000 to 32000, unit specified in Set Engr Unit 1, 1, 7 Engr Unit User set engineering unit Unit specified in Set Engr Unit...
  • Page 42: Appendix 1. Safety Instrumented Systems Installation

    A1.2.2 Diagnostic Response Time Failures that are detected should be reported to The EJX will report an internal failure within 5 Yokogawa. seconds of the fault occurrence. The personnel performing the proof test of the trans- mitter should be trained in SIS operations including A1.2.3 Setup...
  • Page 43 EJX lifetimes beyond 50 years may yield results that A1.2.9 Reliability Data are too optimistic, i.e. the calculated Safety Integrity Level will not be achieved. A detailed Failure Mode, Effects, and Diagnostics Analysis (FMEDA) report is available from Yokogawa with all failure rates and failure modes. IM 01C25T01-01E...
  • Page 44 APPENDIX 1. SAFETY INSTRUMENTED SYSTEMS INSTALLATION A1.2.11 Environmental Limits Safety Assessment The investigation to arrive at a judgment -based on evidence- of The environmental limits of the EJX are specified in the safety achieved by safety- the user’s manual IM 01C25. related systems A1.2.12 Application Limits Further definitions of terms used for safety techniques...
  • Page 45: Revision Record

    REVISION RECORD Title: EJX Series HART Communication Type Manual No.: IM 01C25T01-01E Edition Date Page Revised Item Mar. 2004 – New publication Apr. 2004 – Revise words and phrases 3-26 3.5.16 • Add section 3.5.16 Status Output Oct. 2004 3-27 3.5.17 •...

This manual is also suitable for:

Dpharp ejx series

Table of Contents