Installing The Balance - A&D FX-200i Instruction Manual

Fz-i / fx-i series precision balance
Hide thumbs Also See for FX-200i:
Table of Contents

Advertisement

2-2 Installing the Balance

Install the balance as follows:
1. Place the balance on a solid weighing table. Refer to "3. PRECAUTIONS" for installing the balance.
i
2. The FZ-
series (all) and FX-120
Assemble the pan support, weighing pan and breeze break, on the balance as shown in the
illustration on page 7.
i
i
FX-1200
/2000
Assemble the pan support and weighing pan on the balance as shown in the illustration on page 7.
3. Adjust the leveling feet to level the balance. Confirm it using the bubble spirit level.
4. Confirm that the AC adapter type is correct for the local voltage and power receptacle type.
5. Connect the AC adapter to the balance. Warm up the balance for at least 30 minutes with
nothing on the weighing pan.
Note
The AC adapter plug is protected against dust and may be difficult to insert. When inserting
the plug, turn the plug while pushing on it.
i
i
/200
i
i
/3000
/5000
i
i
/300
/500
8