Class
Item and Parameter
Cp Hi
Upper limit
Cp lo
Lower limit
prt
Data output mode
ap-p
Auto print polarity
ap-b
Auto print difference
dout
Data output
int
Interval time
5-id*
ID number output
5-td
Time/Date output
pU5e
Data output pause
at-f
Auto feed
info
GLP output
ar-d
Zero after output
Refer to "10-10 Comparator Function".
0
Key mode
Auto print mode A
1
(Reference = zero)
Auto print mode B
2
(Reference = last stable value)
3
Stream mode
4
Key mode B (Immediately)
5
Key mode C (When stable)
6
Interval output mode
0
Plus only
1
Minus only
2
Both
0
10 digits
1
100 digits
2
1000 digits
0
Every measurement
1
2 seconds
2
5 seconds
3
10 seconds
4
30 seconds
5
1 minute
2 minute
6
5 minute
7
8
10 minute
0
No output
1
Output
0
No output
1
Time only
2
Date only
3
Time and date
0
No pause
1
Pause (1.6 seconds)
0
Not used
1
Used
0
No output
1
AD-8121 format
2
General data format
0
Not used
1
Used
Factory settings. Digit is a unit of minimum weighing value
Only for the FZ-i series
*
:
38
Description
Accepts the PRINT key only when
the display is stable.
Outputs data when the display is
stable and conditions of ap-p, ap-b
and the reference value are met.
Outputs data continuously.
Accepts the PRINT key regardless
of the display condition.
Accepts the PRINT key immediately
when the display is stable, or waits
for the display to be stable when not.
Uses interval output mode.
Displayed value>Reference
Displayed value<Reference
Regardless of displayed value
Difference between reference value
and displayed value
Interval time for the interval output
mode
(With prt 6)
Selects whether or not the ID
number is output.
Selects the data output interval.
Selects whether or not automatic
feed is performed.
Selects GLP output method.
Adjusts zero automatically after data
is output