6-5 Statistical Calculation Mode (Example of Use)

Here, as an example of use of the statistical calculation mode, mixing of the multiple formulae such
as medicine is described. The mixing process is recorded using the balance and the printer.
In the example, the FX-3000i and the AD-8121B (dump print mode) are connected using the RS-232C
serial interface.
5-5-1 Getting Started
Changing the function table
Changes :
●To enable the statistical calculation mode
●To enable "Zero after output"
Enabling the statistical calculation mode
1 Enter the function table menu.
Press and hold the SAMPLE key until
the function table is displayed, then release the key.
2 Select the application function.
Press the SAMPLE key several times to display
1ap fnc . Then, press the PRINT key to display
1°apf 0 .
3 Change the application function parameter to "2".
Press the RE-ZERO key to display apf 2 . Press the
PRINT key to confirm the change. After
1ba5fnc1
is displayed.
Enabling "Zero after output"
4 Select "Zero after output".
Press the SAMPLE key several times to display 1dout .
Then, press the PRINT key to display
the SAMPLE key several times to display 1°ar-d 0 .
5 Enable "Zero after output".
Press the RE-ZERO key to display ar-d 1 . Then,
press the PRINT key to confirm the change. After
1end , 5if is displayed.
Returning to the weighing mode
6 Press the CAL key to return to the weighing mode.
ba5fnc of
end ,
°prt 0 , and press
23
Press and hold
Press
several times
Press
several times
Press
several times
Return to
the weighing mode