Operation
1 Press the ON:OFF key to turn off the display.
2 While pressing and holding the PRINT key and the SAMPLE key, press the ON:OFF key.
The balance displays
3 Press the PRINT key. Then the balance displays the function switches.
4 Set the switches using the following keys.
SAMPLE key
RE-ZERO key
PRINT key
CAL key

9-2 Initializing the Balance

This function returns the following parameters to factory settings.
Calibration data
Function table
The sample unit mass value (counting mode), 100% reference mass value (percent mode)
External calibration weight
Function switch settings
Statistical data
Internal mass value (Only for the FZ-i series)
Note
Be sure to calibrate the balance after initialization.
Operation
1 Press the ON:OFF key to turn off the display.
2 While pressing and holding the PRINT key and the
SAMPLE key, press the ON:OFF key. The balance
displays
p5 .
3 Press the SAMPLE key to display
4 Press the PRINT key.
To cancel this operation, press the CAL key.
5 Press the RE-ZERO key.
6 Press the PRINT key to initialize the balance.
The balance will automatically return to the weighing
mode.
p5 .
To select the switch to change the parameter. The selected switch blinks.
To change the parameter of the switch selected.
0:To inhibit changes.
To store the new parameter and return to the weighing mode.
To cancel the operation (
weighing mode, press the CAL key once again.
Clr .
35
1:To permit changes
Clr
is displayed.) To return to the
With these keys
held down
Press the key