9. FUNCTION SWITCH AND INITIALIZATION

9-1 Permit or Inhibit

The balance stores parameters that must not be changed unintentionally. There are two switches for
the purpose of protecting these parameters. Each switch can select either "permit" or "inhibit". The
"inhibit" protects parameters against unintentional operations.
Switches (FZ-i series)
Switches (FX-i series)
(The display shown left indicates the factory settings.)
Function table
 0 To inhibit changes to the function table
 1 To permit changes to the function table
Calibration using the internal mass (One-touch calibration)
 0 To inhibit calibration using the internal mass
 1 To permit calibration using the internal mass
Calibration using an external weight
 0 To inhibit calibration using an external weight
 1 To permit calibration using an external weight
Not used
Internal mass value correction
 0 To inhibit correction
 1 To permit correction
(The display shown left indicates the factory settings.)
Function table
 0 To inhibit changes to the function table
 1 To permit changes to the function table
Not used
Calibration using an external weight
 0 To inhibit calibration using an external weight
 1 To permit calibration using an external weight
Not used
Not used