Calibration; Calibration Mode - A&D FX-200i Instruction Manual

Fz-i / fx-i series precision balance
Hide thumbs Also See for FX-200i:
Table of Contents

Advertisement

8-1 Calibration Mode

The balance has the following modes.
Calibration using the internal mass (One-touch calibration, only for the FZ-i series)
Calibration using an external weight
Calibration test using an external weight (Calibration test does not perform calibration.)
Terms
The following terms are defined as follows:
External weight
Calibration weight =
Target weight
Caution
Calibration adjusts the balance for accurate weighing.
Besides periodic calibration and before each use, perform calibration when:
• The balance is installed for the first time.
• The balance has been moved.
• The ambient environment has changed.
Do not allow vibration or drafts to affect the balance during calibration.
To output the data for GLP using the RS-232C serial interface, set "GLP output ( info )" of
"Data output ( dout )". For details, refer to "10. FUNCTION TABLE".
Calibration test is available only when "GLP output ( info )" of "Data output ( dout )" is set to "1" or "2".
When calibrating, use the accessory breeze break. (Models: FZ-i series (all) and FX-120i /200i
/300i /500i)
Caution on using an external weight
The accuracy of an external weight will influence the accuracy of weighing. Select an
appropriate weight as listed below:
Model
i
FX-120
i
FX-200
i
FX-300
i
FX-500
i
FX-1200
i
FX-2000
i
FX-3000
i
FX-5000
The calibration weight in bold type: factory setting
The calibration weight value can be adjusted within the range above.
Display

8. CALIBRATION

=
A weight that you have. Referred to as a calibration weight when used for
calibration.
A weight used for calibration
=
An external weight used for calibration test
Usable calibration weight
100 g, 50 g
200 g, 100 g, 50 g
300 g, 200 g, 100 g, 50 g
500 g, 400 g, 300 g, 200 g, 100 g, 50 g
1000 g, 500 g,
2000 g, 1000 g, 500 g
3000 g, 2000 g, 1000 g, 500 g
5000 g, 4000 g, 3000 g, 2000 g, 1000 g, 500 g
This indicator means "the balance is measuring calibration data".
Do not allow vibration or drafts to affect the balance while this
indicator is displayed.
26
Adjustable range
-0.050 g to +0.050 g
-0.50 g to +0.50 g