Agb1500_Setswitch; Agb1500_Spotcmumeas - Agilent Technologies B1500A User Manual

Semiconductor device analyzer
Hide thumbs Also See for B1500A:
Table of Contents

Advertisement

Syntax
Parameters
Syntax
Parameters
Driver Functions

agb1500_setSwitch

comp
Compliance value (in V or A). It must be voltage for the current
sweep source, or current for the voltage sweep source.
Compliance polarity is automatically set to the same polarity as
the output value, regardless of the specified comp polarity. The
compliance polarity is positive if the output value is 0.
p_comp
Power compliance. 0.001 to 4.0 VA (for MPSMU), or 0.001 to
20.0 VA (for HPSMU) in 0.001 VA resolution. If you enter the
other value, the power compliance is not set.
agb1500_setSwitch
This function sets the output switch of the specified channel.
ViStatus _VI_FUNC agb1500_setSwitch(ViSession vi, ViInt32 channel,
ViInt32 state);
vi
Instrument handle returned from agb1500_init( ).
channel
Slot number of the slot that installs the SMU/MFCMU to be
used. See Table 2-2.
channel=0 specifies all modules.
channel=-1 detects the CMU slot automatically.
state
Output switch setting. 0 (off) or 1 (on).

agb1500_spotCmuMeas

This function executes the high speed spot C measurement, and returns the
measurement data, measurement status, MFCMU monitor data, MFCMU monitor
status, and time stamp data.
Before this function, execute the agb1500_setCmuFreq,
agb1500_forceCmuAcLevel, and agb1500_forceCmuDcBias functions to set the
oscillator and DC bias source.
ViStatus _VI_FUNC agb1500_spotCmuMeas(ViSession vi, ViInt32 channel,
ViInt32 mode, ViReal64 range, ViReal64 value[], ViInt32 status[],
ViReal64 monitor[],ViInt32 status_mon[], ViPReal64 time);
vi
Instrument handle returned from agb1500_init( ).
2-62
Agilent B1500 VXIplug&play Driver User's Guide, Edition 3

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents