Driver Functions

agb1500_setLoadCorrMode

channel
Slot number of the slot that installs the SMU to be used. See
Table 2-2.
mode
Source output mode. 1, 2, 3, 4, -1, -2, -3, or -4. For the log
sweep mode, start and stop must be the same polarity.
1: Voltage-Single-Linear sweep
2: Voltage-Single-Log sweep
3: Voltage-Double-Linear sweep
4: Voltage-Double-Log sweep
-1: Current-Single-Linear sweep (only for SMU)
-2: Current-Single-Log sweep (only for SMU)
-3: Current-Double-Linear sweep (only for SMU)
-4: Current-Double-Log sweep (only for SMU)
range
Output ranging mode. 0 (auto) or positive value (limited auto).
start
Sweep start value (in A or V).
stop
Sweep stop value (in A or V).
point
Number of sweep steps. 1 to 1001.
hold
Hold time. 0 to 655.35 seconds, in 0.01 seconds resolution.
delay
Delay time. 0 to 65.535 seconds, in 0.0001 seconds resolution.
s_delay
Step delay time. 0 to 1.0 seconds, in 0.0001 seconds resolution.
comp
Compliance value (in V or A). It must be voltage for the current
sweep source, or current for the voltage sweep source.
Compliance polarity is automatically set to the same polarity as
the output value, regardless of the specified comp polarity. The
compliance polarity is positive if the output value is 0.
p_comp
Power compliance. Available values are listed below. If you
enter the other value, the power compliance is not set.
0.001 to 4.0 VA (for MPSMU)
0.001 to 20.0 VA (for HPSMU)
Setting resolution: 0.001 VA
agb1500_setLoadCorrMode
This function sets the CMU load correction function ON or OFF.
The agb1500_execLoadCorr function must be executed before this function.
2-52