Agb1500_Setsample - Agilent Technologies B1500A User Manual

Semiconductor device analyzer
Hide thumbs Also See for B1500A:
Table of Contents

Advertisement

Syntax
Parameters
period
Pulse period (in seconds). 0.005 to 5.0 s. 0.1 ms resolution.
If you set period=0, the B1500 automatically sets the pulse
period to 5 msec (for width ≤ 3 ms), width +2 msec (for 3 ms <
width ≤ 100 ms), or width +10 msec (for width > 100 ms).
comp
Compliance value (in V or A). It must be voltage for the current
sweep source, or current for the voltage sweep source.
Compliance polarity is automatically set to the same polarity as
the output value, regardless of the specified comp polarity. The
compliance polarity is positive if the output value is 0.

agb1500_setSample

This function sets the timing parameters of the sampling measurement.
To set the synchronous constant voltage/current source for the sampling
measurement, execute the agb1500_addSampleSyncIv function. To set the linear or
logarithmic sampling mode, execute the agb1500_setSampleMode function. To set
the automatic measurement abort function and the post measurement output value,
execute the agb1500_stopMode function. To define the measurement channels and
start the measurement, execute the agb1500_sampleIv function.
ViStatus _VI_FUNC agb1500_setSample(ViSession vi, ViReal64 bias_hold,
ViReal64 base_hold, ViReal64 interval, ViInt32 point);
vi
Instrument handle returned from agb1500_init( ).
bias_hold
Time since the bias value output until the first sampling point.
Numeric expression. in seconds. 0 (initial setting) to 655.35 s,
resolution 0.01 s.
The following values are also available for interval < 0.002 s.
|bias_hold| will be the time since the sampling start until the
bias value output.
-0.09 to -0.0001 s, resolution 0.0001 s.
base_hold
Hold time of the base value output until the bias value output.
Numeric expression. in seconds. 0 (initial setting) to 655.35 s,
resolution 0.01 s.
Agilent B1500 VXIplug&play Driver User's Guide, Edition 3
period ≥ width +2 msec (for width ≤ 100 ms)
period ≥ width +10 msec (for width > 100 ms)
Driver Functions
agb1500_setSample
2-57