Syntax
Parameters
Driver Functions

agb1500_sweepMiv

agb1500_sweepMiv
This function executes a multi channel sweep measurement by the specified
channels, and returns the number of measurement steps, sweep source data,
measurement data, measurement status, and time stamp data.
Before executing this function, set the sweep source setup by using the
agb1500_setIv function. If you want to use the synchronous sweep source, execute
the agb1500_setSweepSync function.
ViStatus _VI_FUNC agb1500_sweepMiv(ViSession vi, ViInt32 channel[ ],
ViInt32 mode[ ], ViReal64 range[ ], ViPInt32 point, ViReal64 source[ ],
ViReal64 value[ ], ViInt32 status[ ], ViPReal64 time[ ] );
vi
Instrument handle returned from agb1500_init( ).
channel[ ]
Slot number of the slot that installs the SMU to be used. See
Table 2-2.
Enter 0 to the last element of channel[ ]. For example, if you use
two channels, set the array size to 3, specify the channels to the
first and second elements, and enter 0 to the third element.
mode[ ]
Measurement mode. 1 (current) or 2 (voltage).
range[ ]
Measurement ranging mode. 0 (auto), positive value (limited
auto), or negative value (fixed range). For the available values,
see Table 2-3.
point
Number of measurement steps. Returned data.
source[ ]
Sweep source setup data. Returned data. To disable the source
data output, set 0 (NULL pointer) instead of array.
value[ ]
Measurement data. Returned data.
status[ ]
Measurement status. Returned data. See "Status Code" on page
2-14. To disable the status data output, set 0 (NULL pointer)
instead of array.
time[ ]
Time stamp data (measurement start time). Returned data. To
disable the time stamp data output, set 0 (NULL pointer)
instead of array.
2-72
Agilent B1500 VXIplug&play Driver User's Guide, Edition 3