Serial Interface (continue)
Item
f46
DP / Delimiter
f47
Baud rate
f48
Data bit, parity
*: Initial settings.
Description of "Stream Mode"
f37 0, f40 0
Object
Operation
Data is output in every sampling (when refreshing the display).
Use this mode to output data to an external display (Data may not be output
due to timing of the baud rate and internal sampling rate). If data is printed
with pressing the
Description of "Manual Print"
f37 1, f40 1
Object
Operation
When pressing the
Description of "Auto Print Mode"
f37 2, f37 3, f40 2 or f40 3
Object
Operation
When the weight data varies from the "inhibit region for output" to the
"permission region of output", the stable data is output just once. If you use
this mode, set f02 except f02 0.
For weighing (and removing) each object and printing the data.
In case of f37 2, f40 2
"Inhibit region for output"
In case of f37 3, f40 3
-5d
"Permission region of output" < -5d, +5d < "permission region of output".
Description of "Delay
Description of "Delay
Description of "Delay for continual data"
Description of "Delay
Object
F38, F42
Operation
This function can be used in the "Auto print mode" and "Manual print mode".
When using a non-buffered printer, set to f42 3 and f43 1 (or f38 1).
AD-4407 Weighing Indicator
Parameter
* 0
DP:point(.) / Delimiter:comma(,)
1
DP:comma(,)/ Delimiter:semicolon(;)
0
600 bps
1
1200 bps
** 2
2400 bps
3
4800 bps
4
9600 bps
** 0
Data 7bits, Even parity
1
Data 7bits, Odd parity
2
Data 8bits, Non parity
bps:
key on the printer, use the stream mode.
PRINT
PRINT
"inhibit region for output"
for continual data"
for continual data"
for continual data"
Description
bit per second.
key, the stable weighing data is output just once.
+5d. +5d < "permission region of output".
+5d.
d: division (weighing interval) of the first range.