Agilent Technologies 3070 System Installation Manual page 37

Board test systems
Hide thumbs Also See for 3070:
Table of Contents

Advertisement

Chapter 1: MS Windows System Installation Procedure: Setting-Up Other Hardware
© Agilent Technologies 2001–2003
4 Use a 5-millimeter hex key wrench to remove the
safety shroud from around the testhead's pin field.
5 After installing the automation equipment on the
testhead, reach under the testhead and disconnect
the plug from the Auxiliary connector (J3) on the
System Card
(Figure
1-13).
Figure 1-13 Connecting the EMO cable
6 Remove the jumper from terminals 11 and 12 on the
plug and connect the Emergency Shutdown (EMO)
switch on the automation equipment to pins 11 and
12 on the plug. Then reinstall the plug on the System
Card.
Agilent 3070 System Installation Manual (MS Windows Version)
WARNING
DO NOT operate the system without an EMO
switch properly wired and accessible to the
operator. Failure to provide EMO capability may
result in death or serious bodily injury.
7 Boot the testhead. When the testhead is booted, press
the EMO switch on the automation equipment to
verify that it is functioning properly.
1-31