Chapter 1: MS Windows System Installation Procedure: Setting Up the KVM and Support Arms
© Agilent Technologies 2001–2003
Table 1-2
Install the KVM and support arms (continued)
Task
6 Route the KVM cables:
Agilent 3070 System Installation Manual (MS Windows Version)
Step
CAUTION
If the KVM cables are not routed with enough slack, they can be
damaged when the support arms are adjusted.
describes this further.
NOTE
The keyboard and mouse is shipped with long enough cables to
accommodate installation on the left side.They are folded back on
themselves in the testhead cable trough (identified in
page 1-11).
For 307X and 317X systems, a second video monitor cable is
pre-installed to accommodate KVM installation on the left side.
For 327X systems, the existing video monitor cable is long enough to
reach to the left side. Re-route it through the testhead cable troughs. Cut
cable ties inside the pod, if necessary.
It may be necessary to obtain a 8120-1763 power cord extension cable to
connect power from the PDU to the video monitor. Route the power
cable through the testhead cable trough.
• Route the KVM cables through the support arm cable troughs.
Figure 1-4
on page 1-14
Figure 1-1
on
1-9