Chapter 1: MS Windows System Installation Procedure: Verifying the Testhead
Verifying the
Testhead
© Agilent Technologies 2001–2003

Run AutoAdjust All

CAUTION
Allow the testhead to warm up for 30 minutes
before running AutoAdjust All.
If the testhead temperature is not allowed to
stabilize, diagnostic and testing failures may
appear later.
1 Verify a
logon.
service3070
2 Start Diagnostics if it does not start automatically:
a Click Start > Programs > Agilent 3070 > Korn
Shell.
b At the Korn Shell window prompt, enter:
dgn
3 From the Service Package - Level 1 menu, select
Config.
4 Select DGN Config.
5 Press function key F6 in the AutoAdjust menu to run
AutoAdjust All.
6 Verify that there are no errors.
Agilent 3070 System Installation Manual (MS Windows Version)

Run Full Diagnostics

NOTE
Compressed air must be attached to the system to
run full Diagnostics.
1 Place a pin verification fixture on the testhead, then
lock it down.
2 Change Manual Intervention to Yes (press Next Value).
NOTE
Do not press Save Config as that will make the
Manual Intervention selection permanent.
3 Run Full Diagnostics.
4 Verify that there are no errors.
5 Logout of DGN.
1-39