Performance Tests (Voltage, Resistance, And Frequency) - Fluke 2620A User Manual

Hydra series ii data acquisition unit, hydra series ii data logger
Hide thumbs Also See for 2620A:
Table of Contents

Advertisement

2620A, 2625A
Users Manual
Function
DC Volts
Voltages greater than 150V can only be applied to channels 0, 1, and 11. The rear Input Module must be
installed when measuring ac volts on channel 0.
For 2-terminal measurements, the resistance accuracy given in this table applies to Channel 0 and makes
allowance for up to 0.05 ohm of lead wire resistance. You must add any additional lead wire resistance
present in your set up to the resistance values given in this table.
Resistance
6-8
Table 6-3. Performance Tests (Voltage, Resistance, and Frequency)
Range
Input Level
300 mV
300 mV
300 mV
3V
-3V
30V
150V
300V
Voltages greater than 150V can only be applied to channels 0,1, and 11.
300 mV
300 mV
300 mV
3V
3V
30V
150V
300V
Using inputs in decades of 3:
300
short
300
3 k
short
3 k
30 k
30 k
300 k
300 k
3 M
3 M
Frequency
0V
--
150 mV
---
290 mV
---
2.9V
---
-2.9V
---
29V
---
150V
---
290V
---
Note
20 mV
1 kHz
20 mV
100 kHz
290 mV
1 kHz
290 mV
100 kHz
2.9V
1 kHz
29V
1 kHz
150V
1 kHz
290V
1 kHz
Note
Note
Display Accuracy
(1 year, 18-28 C)
MIN
MAX
-0.02
0.02
149.94
150.06
289.91
290.09
2.8991
2.9009
-2.9009
-2.8991
28.991
29.009
149.94
150.06
289.91
290.09
19.71
20.29
18.50
21.50
289.28
290.72
275.00
305.00
2.8937
2.9063
28.931
29.069
149.54
150.46
289.34
290.66
0.00
0.09
299.90
300.15
0.0000
0.0003
2.9989
3.0011
29.990
30.010
299.88
300.12
2.9979
3.0021