Setting Alarms; Alarm Limits; Rtd Temperature - Fluke 2620A User Manual

Hydra series ii data acquisition unit, hydra series ii data logger
Hide thumbs Also See for 2620A:
Table of Contents

Advertisement

Channel
G
PRESS
THESE
D
BUTTONS:
TO
0
SELECT
1
FROM
.
THESE
.
CHOICES:
20
Note 1. Pt selects RTD temperature measurement (DIN/IEC 751). See Table 3-5 for J, K, E, T, N, R, S, b,
and C thermocouple selections.
Note 2. 4T not available on channels 0 and 11 through 20.
Note 3. R0 default is 100.00 . A unique R0 value can be set for each channel.
Temperature units can be displayed in degrees Celsius (ºC) or Fahrenheit
(ºF). To switch this setting between ºC and ºF, start with the instrument
powered off, then press and hold
setting can also be changed through the Computer Interface with the
TEMP_CONFIG command (refer to Chapter 4.)
RTD temperature measurement uses a resistance-temperature detector (RTD). RTDs,
while usually larger and more expensive than thermocouples, are frequently used where
accuracy, stability, and repeatability are important. The resistance of an RTD varies
directly with the sensor temperature. Passing a current through this resistance generates a
proportional voltage that can be accurately translated into a temperature reading. The
instrument supports the DIN/IEC 751 RTD type.

Setting Alarms

Alarm Limits

If you press
Therefore, for a new channel, use
function before selecting
Two alarm limits ( and ) can be defined for each analog input channel. An alarm
occurs when the measured value on the channel moves above the HI value or below the
LO value. With the desired channel already selected from Inactive Mode, verify or
change these limits using the procedure shown in Table 3-7. If necessary, refer to
"Entering and Changing Numeric Values" for a more detailed description of the number
changing technique used here.
Table 3-6. RTD Temperature
Function
(Note 1)
G
F
E
D
OFF
V DC
V AC
Hz
C or F
Note
Note
for a channel that is OFF, an error beep will result.
.
Operating the Instrument from the Front Panel
Type
Terminals
(Note 2)
G
G
E
D
D
J
2T
4T
K
E
T
N
R
S
b
C
Pt
while pressing POWER ON. The
to define the channel's measurement
Channel Configuration
R0
(Ice Point)
(Note 3)
G
E
D
E
H
J
100.00 Completes
selection
and returns
to Inactive
Mode
3
3-9